Ventricular tachycardia rate and morphology determine energy and current requirements for transthoracic cardioversion.

نویسندگان

  • R E Kerber
  • M G Kienzle
  • B Olshansky
  • A L Waldo
  • D Wilber
  • M D Carlson
  • A M Aschoff
  • S Birger
  • L Fugatt
  • S Walsh
چکیده

BACKGROUND The electrical current and energy required to terminate ventricular tachyarrhythmias are known to vary by arrhythmia: Ventricular tachycardia (VT) is generally considered to require less energy than ventricular fibrillation (VF). The hypothesis of our study was that current requirements for transthoracic termination of VT are further determined by VT rate and QRS complex morphology. METHODS AND RESULTS We prospectively studied 203 patients who received a total of 569 shocks for VT or VF by following a current-based protocol. This protocol recommended shocks for VT beginning at 18 A (70 +/- 22 J) and shocks for VF beginning at 25 or 30 A (137 +/- 52 J or 221 +/- 70 J). The ventricular tachyarrhythmias were subclassified as monomorphic VT (MVT): uniform QRS complex morphology on surface electrocardiogram and heart rate greater than 100 beats per minute; polymorphic VT (PVT): nonuniform QRS complex morphology and heart rate less than or equal to 300 beats per minute; or VF: nonuniform QRS complex morphology and heart rate greater than 300 beats per minute. We found that shocks of 18 A and 25 A for terminating MVT had success rates of 69% and 82%, respectively, whereas such low-current shocks were less successful for PVT (33% at 18 A) and for VF (19% at 18 A, 53% at 25 A). High-current shocks of 35 A and 40 A were equally successful for the three ventricular tachyarrhythmias. Subdividing MVT revealed that slower MVT (heart rate less than 200 beats per minute) had a significantly better success rate with low-current shocks of 18 A and 25 A than did faster MVT (greater than 200 beats per minute) (89% versus 72% success, p less than 0.01). Bundle branch block morphology, QRS axis, and duration of ventricular tachyarrhythmia did not alter current requirements. CONCLUSIONS Heart rate and electrocardiographic degree of organization of ventricular tachycardia are important determinants of transthoracic energy and current requirements for cardioversion and defibrillation. Transthoracic termination of MVT requires relatively low current or energy, but PVT behaves more like VF and requires higher electrical current or energy.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Cardioversion and the digitalized patient

Transthoracic synchronized direct current discharge or cardioversion was introduced nearly 25 years ago to restore sinus rhythm in patients with diverse tachyarrhythmias (I). This simple, expeditious and nearly complication-freemethod has withstood the test of time. The likelihood of serious adverse reactions is related more to the underlying mechanism being treated than to the electrical disch...

متن کامل

Safety and efficacy of ibutilide in cardioversion of atrial flutter and fibrillation.

This article reviews the safety and efficacy of ibutilide for use in patients with atrial fibrillation and flutter. Ibutilide, a class III antiarrhythmic agent, is primarily used for conversion of atrial flutter and fibrillation and is a good alternative to electrical cardioversion. Ibutilide has a conversion rate of up to 75% to 80% in recent-onset atrial fibrillation and flutter; the conversi...

متن کامل

Ventricular tachycardia initiated by high energy cardioversion in a patient with an implantable cardioverter defibrillator.

A transvenous implantable cardioverter defibrillator (ICD) was implanted into a 58 year old woman with idiopathic dilated cardiomyopathy who had drug refractory monomorphic ventricular tachycardia (VT). Antitachycardia pacing failed to terminate the VT; termination was attempted at 24 J, which was above the defibrillation threshold. When cardioversion at 24 J was delivered, VT with a different ...

متن کامل

Measurement of cardioversion/defibrillation thresholds in man by a truncated exponential waveform and an apical patch-superior vena caval spring electrode configuration.

Defibrillation/cardioversion thresholds were measured in 33 patients undergoing defibrillator implants. Each patient had a 12 cm2 patch placed near the left ventricular apex via a thoracotomy and a 10 cm2 spring lead placed pervenously at the right atrial-superior vena caval junction. Ventricular tachycardia of stable morphology, polymorphic ventricular tachycardia, or ventricular fibrillation ...

متن کامل

Comparative efficacy of transvenous cardioversion and pacing in patients with sustained ventricular tachycardia: a prospective, randomized, crossover study.

We performed a prospective, randomized crossover study to evaluate the comparative efficacy of transvenous cardioversion and rapid ventricular pacing for termination of induced ventricular tachycardia in patients with spontaneous ventricular tachycardia and organic heart disease. Sixty-two episodes of ventricular tachycardia were induced in 15 patients, mean age 60 +/- 10 years, during electrop...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Circulation

دوره 85 1  شماره 

صفحات  -