Distribution and Repeatability of Corneal Astigmatism Measurements (Magnitude and Axis) Evaluated With Color Light Emitting Diode Reflection Topography

نویسندگان

  • Anastasios John Kanellopoulos
  • George Asimellis
چکیده

PURPOSE To evaluate and investigate the distribution and repeatability of anterior corneal surface astigmatism measurements (axis and magnitude) using a novel corneal topographer. METHODS Anterior corneal surface astigmatism was investigated in a total of 195 eyes using a novel multicolored spot reflection topographer (Cassini; i-Optics). Two patient groups were studied, a younger-age group A and an older-age group B. Three consecutive acquisitions were obtained from each eye. The repeatability of measurement was assessed using Bland-Altman plot analysis and is reported as the coefficient of repeatability. RESULTS Group A (average age 34.3 years) had on average with-the-rule astigmatism, whereas the older-age group B (average age 72.3 years) had on average against-the-rule astigmatism. Average astigmatism magnitude measurement repeatability in group A was 0.4 diopters (D) and in group B 0.4 D. Average astigmatism axis measurement repeatability in group A was 5.4 degrees and in group B 5.5 degrees. The axis measurement repeatability improved with increasing magnitude of astigmatism: in the subgroups with astigmatism between 3.0 and 6.0 D, the axis repeatability was 1.4 degrees (group A) and 1.2 degrees (group B), whereas in the subgroups with astigmatism larger than 6.0 D, the repeatability was 1.1 and 0.6 degrees, respectively. CONCLUSIONS This novel corneal topography device seems to offer high precision in reporting corneal astigmatism. This study reaffirms the established trend of a corneal astigmatism shift from an average "with-the-rule" to "against-the-rule" with aging.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Color light-emitting diode reflection topography: validation of keratometric repeatability in a large sample of wide cylindrical-range corneas

PURPOSE To investigate repeatability of steep and flat keratometry measurements, as well as astigmatism axis in cohorts with normal range and regular astigmatic such as: eyes following laser-assisted in situ keratomileusis (LASIK) and normal population, as well as cohorts of high and irregular astigmatism such as keratoconic eyes, and keratoconic eyes following corneal collagen cross-linking, e...

متن کامل

Investigation of a real-time location system of corneal astigmatic axis

BACKGROUND To construct a real-time computerized location system (RCLS) to analyze and display the axis of corneal astigmatism and to compare its accuracy with the Scheimpflug method. METHODS Fifty-seven eyes of 39 volunteers with corneal astigmatism more than 1.00 diopter (D) were recruited. The RCLS was composed of a circular light-emitting diode (LED) light source, surgical microscope, sur...

متن کامل

Comparability and repeatability of different methods of corneal astigmatism assessment

Purpose To assess the comparability and repeatability of keratometric and astigmatism values measured by four techniques: Orbscan IIz® (Bausch and Lomb), Lenstar LS 900® (Haag-Streit), Cassini® (i-Optics), and Total Cassini (anterior + posterior surface), in healthy volunteers. Patients and methods Fifteen healthy volunteers (30 eyes) were assessed by the four techniques. In each eye, three c...

متن کامل

Comparative evaluation of the effect of Light Emitting Diode (LED) and Quartz Tungsten Halogen (QTH) light curing units on color stability of Filtek Z350 XT

Introduction:Discoloration of the resin-based composites is a common problem in restorative dentistry. There are many factors associated with the discoloration of dental materials in the oral environment. The purpose of this study was to evaluate the color changes in a nano-composite cured with a quartz-tungsten-halogen (QTH) and light emitting diode (LED) unit. Methods:80 disk-shaped specim...

متن کامل

Comparison of keratometric and topographic cylinder and axis measurements on normal corneas with low astigmatism.

PURPOSE To evaluate agreement in measurements of astigmatic axis power and location between keratometry and computer assisted videokeratography (corneal topography) on normal corneas with less than 1.50 D of idiopathic astigmatism. METHODS Keratometric readings with the 10 SL/O Zeiss ophthalmometer and corneal topographic maps with the TMS-1 were obtained by two independent examiners on 32 no...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 34  شماره 

صفحات  -

تاریخ انتشار 2015