1.0 s Ultrafast MRI in non-sedated infants after reduction with spica casting for developmental dysplasia of the hip: a feasibility study
نویسندگان
چکیده
PURPOSE The aim of this study was to first develop and use 1.0 s ultrafast magnetic resonance imaging (MRI) to confirm the location of the femoral head in non-sedated infants with developmental dysplasia of the hip (DDH) after reduction with spica cast application in clinical settings. METHODS The ultrafast acquisition was achieved by employing a balanced steady-state free precession sequence and immobilizing the patient with dedicated sandbags. On completion of the ultrafast MRI study, all infants were sedated for conventional MRI scanning. Two orthopaedic surgeons retrospectively evaluated the image quality, result of the reduction and total MRI study time (including patient immobilization, coil setup, and scanning) in 14 DDHs of 13 infants (one with bilateral DDHs). RESULTS Both reviewers stated that there were no motion artefacts for non-sedated infants during the ultrafast MRI and that the quality of both the ultrafast and conventional MRI images were acceptable to assess the femoral head location. Assessment of the reduction procedure resulted in two hips being categorized as 'incomplete reduction' requiring a re-reduction procedure. The total study time of ultrafast and conventional MRI was 6 ± 1 min and 14 ± 3 min, respectively (P < 0.001). No complications due to sedation, such as hypoxia, were reported. The average sedation waiting time was 1 h 25 min ± 34 min. CONCLUSION The ultrafast MRI procedure reported here can be readily employed to confirm the location of the femoral head in infants with DDHs, without the use of any sedation.
منابع مشابه
Popliteal angle as an indicator for successful closed reduction of developmental dysplasia of the hip.
PURPOSE To evaluate the use of the popliteal angle as an indicator for successful closed reduction of developmental dysplasia of the hip (DDH) in children with delayed presentation. METHODS 29 patients aged 6 to 18 months underwent successful closed reduction for unilateral DDH under general anaesthesia. Using a graduated goniometer, the popliteal angle was measured with the hip abducted with...
متن کاملSpica magnetic resonance imaging for determination of abduction angle: initial results and reproducibility assessment
PURPOSE Spica magnectic resonance imaging (MRI) is an established technique for postoperative determination of hip reduction in patients treated for developmental dysplasia of the hip (DDH). A hip abduction angle >55° is considered excessive and has been associated with epiphyseal osteonecrosis. Our purpose was to establish objective criteria for measuring hip abduction angles on MRI after hip ...
متن کاملSemirigid Abduction Bracing is Effective Treatment of Reducible Developmental Dysplastic Hips after Failure of Pavlik Harness
Developmental dysplasia of the hip [DDH] is a common condition of the childhood, affecting 1.6 to 28.5 newborns per 1000 live births [1]. Treatment is age dependent, progressing from abduction splinting with nonrigid or semirigid braces to closed reduction with hip spica cast and finally to open surgical procedures. Pavlik harness is a nonrigid brace widely accepted to treat infants younger tha...
متن کاملPhysical and Clinical Evaluation of Hip Spica Cast applied with Three-slab Technique using Fibreglass Material
Introduction: Hip spica casting is an important component of treatment for developmental dysplasia of the hip (DDH) and popular treatment method for femur fractures in children. Breakage at the hip region is a relatively common problem of this cast. We have developed a three-slab technique of hip spica application using fibreglass as the cast material. The purpose of this review was to evaluate...
متن کاملAn improved method for measuring hip abduction in spica after surgical reduction for developmental dysplasia of the hip
PURPOSE Excessive in-spica abduction is a risk factor for oste-onecrosis after surgical reduction for developmental dysplasia of the hip (DDH). The traditional method for radiographically measuring hip abduction using axial imaging does not reflect the true angle, which usually lies in an oblique plane. The purpose of this study was to describe a novel method for measuring true hip position usi...
متن کامل