Short term in vivo precision of whole body composition measurements on the Horizon A densitometer.
نویسندگان
چکیده
INTRODUCTION Dual-energy X-ray absorptiometry (DXA) is increasingly being used to assess abnormalities in body composition associated with a wide variety of conditions including obesity, sarcopenia, diabetes, anorexia, human immunodeficiency virus lipodystrophy, malabsorption and neuromuscular disorders. The objective of this study was to determine the short-term in vivo precision and least significant change in serial body composition measurements provided by the Hologic Horizon A densitometer as there is no published, peer reviewed body composition precision data on Horizon scanners in the medical literature. METHODS After acquiring two new Hologic Horizon A densitometers, short term precision scanning was performed on the two scanners using different groups of 30 volunteers, who had duplicate scans with all scans performed by the same operator. Using the International Society of Clinical Densitometry's advanced precision tool, the scan data were recorded and the coefficient of variation (CV) and least significant change (LSC) determined for total fat, lean and bone mineral content (BMC) and bone mineral density (BMD). RESULTS Coefficients of variation of 0.78% and 0.77% were achieved for total fat (LSC at 95% confidence 2.15%) and 0.52% and 0.40% for total lean (LSC at 95% confidence 1.45% and 1.11%). CONCLUSIONS These results indicate that the two Horizon A scanners in this study exceeded the minimum acceptable precision values by roughly a factor of four and are therefore capable of detecting small differences in lean and fat mass of 1-2% between two measurements at 95% confidence.
منابع مشابه
Precision of regional bone mineral measurements obtained from total-body scans.
Newer bone densitometers using dual-photon absorptiometry (DPA) or dual energy x-ray absorptiometry (DEXA) are capable of measuring the total-body bone mineral; regional analysis of these scans would have significant utility if adequate precision were possible. This study investigated short term precision by weekly scanning (three to five times) normal subjects (total 48 scans) and long term pr...
متن کاملEffects of Short-Term Interval Training Courses on Fitness and Weight Loss of Untrained Girls
Due to the increase in obesity and sedentary lifestyle in today's world, encouraging the individuals to do appropriate and academic physical activities and studying sports programs with different structures are if great importance. The present paper aims to study the effect of 30 minutes interval short courses of running (4 rounds × 7:30 minutes) with a mean intensity of 60-75% of heart rate ...
متن کاملEffect of sub lethal concentration of imidacloprid on proximate body composition of Labeo rohita
Randomly selected 320 fingerling of Labeo rohita were collected from Faheem fish farm, Mattital Road, Multan. Groups of 40 fish each were exposed to subleathal concentration of imidacloprid (120 mg L-1) under short (2, 4 and 8 days) and long term (16, 32 and 64 days) conditions to access the effect of this most extensively used pesticide on the body composition parameters of economically most i...
متن کاملEffect of sub lethal concentration of imidacloprid on proximate body composition of Labeo rohita
Randomly selected 320 fingerling of Labeo rohita were collected from Faheem fish farm, Mattital Road, Multan. Groups of 40 fish each were exposed to subleathal concentration of imidacloprid (120 mg L-1) under short (2, 4 and 8 days) and long term (16, 32 and 64 days) conditions to access the effect of this most extensively used pesticide on the body composition parameters of economically most i...
متن کاملMonitoring BMD with DXA: Short- and Long-term Precision
The ability to monitor patient response to osteoporosis therapy with DXA depends on long-term in vivo precision and expected treatment response to therapy at a particular region. A review of recent peer-reviewed literature (57 studies) of short-term in vivo coefficient of variations (CV’s) found a statistically significant difference between manufacturers for the AP spine and femoral neck, but ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of medical imaging and radiation oncology
دوره 62 2 شماره
صفحات -
تاریخ انتشار 2018