Comparison of Wet Radiofrequency Ablation with Dry Radiofrequency Ablation and Radiofrequency Ablation Using Hypertonic Saline Preinjection: Ex Vivo Bovine Liver
نویسندگان
چکیده
OBJECTIVE We wished to compare the in-vitro efficiency of wet radiofrequency(RF) ablation with the efficiency of dry RF ablation and RF ablation with preinjection of NaCl solutions using excised bovine liver. MATERIALS AND METHODS Radiofrequency was applied to excised bovine livers in a monopolar mode for 10 minutes using a 200 W generator and a perfused-cooled electrode with or without injection or slow infusion of NaCl solutions. After placing the perfused-cooled electrode in the explanted liver, 50 ablation zones were created with five different regimens: group A; standard dry RF ablation, group B; RF ablation with 11 mL of 5% NaCl solution preinjection, group C; RF ablation with infusion of 11 mL of 5% NaCl solution at a rate of 1 mL/min, group D; RFA with 6 mL of 36% NaCl solution preinjection, group E; RF ablation with infusion of 6 mL of 36% NaCl solution at a rate of 0.5 mL/min. In groups C and E, infusion of the NaCl solutions was started 1 min before RF ablation and then maintained during RF ablation (wet RF ablation). During RF ablation, we measured the tissue temperature at 15 mm from the electrode. The dimensions of the ablation zones and changes in impedance, current and liver temperature during RF ablation were then compared between the groups. RESULTS With injection or infusion of NaCl solutions, the mean initial tissue impedance prior to RF ablation was significantly less in groups B, C, D, and E (43-75 ohm) than for group A (80 ohm) (p < 0.05). During RF ablation, the tissue impedance was well controlled in groups C and E, but it was often rapidly increased to more than 200 ohm in groups A and B. In group D, the impedance was well controlled in six of ten trials but it was increased in four trials (40%) 7 min after starting RF ablation. As consequences, the mean current was higher for groups C, D, and E than for the other groups: 401+/-145 mA in group A, 287+/-32 mA in group B, 1907+/-96 mA in group C, 1649+/-514 mA in group D, and 1968+/-108 mA in group E (p< 0.05). In addition, the volumes of RF-induced coagulation necrosis were greater in groups C and E than in group D, which was greater than in groups A and B than in group E (p < 0.05); 14.3+/-3.0 cm(3) in group A; 12.4+/-3.8 cm(3) in group B; 80.9+/-9.9 cm(3) in group C; 45.3+/-11.3 cm(3) in group D and 81.6+/-8.6 cm(3) in group E. The tissue temperature measured at 15 mm from the electrode was higher in groups C, D and E than other groups (p < 0.05): 53+/-12 degrees C in group A, 42+/-2 degrees C in group B, 93+/-8 degrees C in group C; 79+/-12 degrees C in group D and 83+/-8 degrees C in group E. CONCLUSION Wet RF ablation with 5% or 36% NaCl solutions shows better efficiency in creating a large ablation zone than does dry RF ablation or RF ablation with preinjection of NaCl solutions.
منابع مشابه
An Investigation of Brightness Changes in Ultrasound Images due to Temperature Variations in Liver Tissue during Radiofrequency Ablation
Introduction: one of the minimally invasive methods of treatmenting liver malignancies is Radiofrequency Ablation (RFA) which can be applied to primary or secondary tumors. Nevertheless, the disease recurrence is probable after RFA application partially due to the inadequate capability of temperature monitoring of the target tissue and visualizing the thermal damage. The goal of this work was t...
متن کاملOptimization of Wet Radiofrequency Ablation Using a Perfused-Cooled Electrode: A Comparative Study in Ex Vivo Bovine Livers
OBJECTIVE To determine the optimized protocol for wet monopolar radiofrequency ablation (RFA) using a perfused-cooled electrode to induce coagulation necrosis in the ex vivo bovine liver. MATERIALS AND METHODS Radiofrequency was applied to excised bovine livers in a monopolar mode using a 200W generator with an internally cooled electrode (groups A and B) or a perfused-cooled electrode (group...
متن کاملMoving-Shot versus Fixed Electrode Techniques for Radiofrequency Ablation: Comparison in an Ex-Vivo Bovine Liver Tissue Model
OBJECTIVE To compare the ablation characteristics of the moving-shot technique (MST) and the fixed electrode technique (FET) for radiofrequency (RF) ablation in an ex-vivo bovine liver tissue model. MATERIALS AND METHODS We performed RF ablation using FET in 110 bovine liver blocks using 11 different ablation times ranging from 5 seconds to 5 minutes (10 blocks per each time duration). Ten bo...
متن کاملHepatic Radiofrequency Ablation Using Multiple Probes: Ex Vivo and In Vivo Comparative Studies of Monopolar versus Multipolar Modes
OBJECTIVE We wanted to compare the efficiency of multipolar radiofrequency ablation (RFA) using three perfused-cooled electrodes with multiple overlapping and simultaneous monopolar techniques for creating an ablation zone in ex vivo bovine livers and in in vivo porcine livers. MATERIALS AND METHODS In the ex vivo experiments, we used a 200 W generator (Valleylab, CC-3 model) and three perfus...
متن کاملPerfused hypertonic-saline-augmented needle enlarges ablation zones in ex vivo porcine livers
There is a great clinical requirement to improve radiofrequency ablation (RFA) efficacy and create larger coagulation necrotic areas. The aim of the present study was to assess the ability of a hypertonic-saline (HS)-enhanced multipolar RFA technique using a perfused electrode to increase RF-created coagulation necrosis, and to compare that technique with natural saline-augmented needle and con...
متن کامل