Enhancing the sensitivity of a micro-diaphragm resonating sensor by effectively positioning the mass on the membrane
نویسندگان
چکیده
The detection of biomarkers in the liquid phase using mechanical sensors is difficult because of noise caused by the liquid. To reduce and verify the side effects of liquid loading, we performed calculations and experiments to determine the shift in resonant frequency according to the loading conditions. A 2-μm-thick piezoelectric rectangular micro-diaphragm with a 500 × 500 μm membrane was used. These dimensions were determined such that there would be an analogous resonant frequency shift ratio in both (1, 1) and (2, 2) modes. By calculating and measuring the resonant frequency, we verified that the resonant frequency of the sensor would change only through contact with the liquid, even the resonant frequency change by only liquid much higher than the changes caused by the nanoparticles. The real signal constituted only 0.017% of the initial resonant frequency. To enhance the sensitivity by reducing the unexpected surface stress in the liquid, the liquid was dropped onto the surface of the micro-diaphragm. This resulted in an improvement of more than 10 times the sensitivity in both modes. In addition, by controlling the position in the micro-diaphragm resonating sensor, more sensitive positions with large displacements were determined according to each mode.
منابع مشابه
Design of High Sensitivity and Linearity Microelectromechanical Systems Capacitive Tire Pressure Sensor using Stepped Membrane
This paper is focused on a novel design of stepped diaphragm for MEMS capacitive pressure sensor used in tire pressure monitoring system. The structure of sensor diaphragm plays a key role for determining the sensitivity of the sensor and the non-linearity of the output.First the structures of two capacitive pressure sensors with clamped square flatdiaphragms, with different thicknesses are inv...
متن کاملSimulation and Modeling of a High Sensitivity Micro-electro-mechanical Systems Capacitive Pressure Sensor with Small Size and Clamped Square Diaphragm
This paper proposes a Micro-electro-mechanical (MEMS) capacitive pressure sensor that relies on the movable electrode displaced like a flat plate equal to the maximum center deflection of diaphragm. The diaphragm, movable electrode and mechanical coupling are made of polysilicon, gold and Si3N4, respectively. The fixed electrode is gold and the substrate is Pyrex glass. This proposed method inc...
متن کاملA mechanical micro molecular mass sensor
One of the bio-sensing mechanisms is mechanical. Rather than measuring shift in resonance frequency, we adopt to measure the change in spring constant due to adsorption, as one of the fundamental sensing mechanism. This study explains determination of spring constant of a surface functionalized micro machined micro cantilever, which resonates in a trapezoidal cavity-on Silicon wafer, with the ...
متن کاملA mechanical micro molecular mass sensor
One of the bio-sensing mechanisms is mechanical. Rather than measuring shift in resonance frequency, we adopt to measure the change in spring constant due to adsorption, as one of the fundamental sensing mechanism. This study explains determination of spring constant of a surface functionalized micro machined micro cantilever, which resonates in a trapezoidal cavity-on Silicon wafer, with the ...
متن کاملDesign and Simulation of a Fluidic Micro-Bio-Sensor Based on Resonator Array
In this paper, a fluidic biosensor with possibility to fabricate by Micro-Electro-Mechanical Systems (MEMS) technology is proposed for biomedical mass detection and lab-on-chip applications. This is designed by electromechanical coupling of harmonic micromechanical resonators with harmonic springers as a mechanical resonator array. It can disperse mechanical wave along the array by electrostati...
متن کامل