Improvement of POC-CCA Interpretation by Using Lyophilization of Urine from Patients with Schistosoma mansoni Low Worm Burden: Towards an Elimination of Doubts about the Concept of Trace

نویسندگان

  • Paulo Marcos Zech Coelho
  • Liliane Maria Vidal Siqueira
  • Rafaella Fortini Queiroz Grenfell
  • Nathalie Bonatti Franco Almeida
  • Naftale Katz
  • Áureo Almeida
  • Nídia Francisca de Figueiredo Carneiro
  • Edward Oliveira
چکیده

BACKGROUND Accurate diagnostic techniques for schistosomiasis are essential for prevalence determination and identification of positive patients. A point-of-care test for detecting schistosome circulating cathodic antigen (POC-CCA) has been evaluated for its accuracy in different endemic regions. This reagent strip/dipstick based assay has showed high sensitivity for individuals with high or moderate worm burden, but the interpretation of light infections is less clear, especially for trace readings. METHODOLOGY/PRINCIPAL FINDINGS We introduced a urine lyophilization step to the POC-CCA assay to improve its sensitivity and clarify the interpretation of traces. We evaluated POC-CCA sensitivity and specificity within individuals with low parasite burdens in a Brazilian endemic area where a high number of traces were detected. Patients that were positive for other helminths were also evaluated for cross reactions. In all cases, a combined parasitological diagnosis using Kato-Katz (24 slides) and Saline Gradient (1 g of feces) were used as reference. At baseline, diagnosis by POC-CCA (1-2 cassettes) showed 6% sensitivity, inaccurately predicting a low prevalence of Schistosoma mansoni infections (2 POC-CCA positives/32 egg positives). After urine lyophilization, the sensitivity was increased significantly (p < 0.05). Prevalence rates changed from 2% to 32% (27 POC-CCA positives/32 egg positives), equivalent to parasitological techniques. Most of the trace readings changed to positive after lyophilization while some negatives turned into traces. Cross reaction analysis confirmed the specificity of POC-CCA. CONCLUSIONS/SIGNIFICANCE Trace readings cannot be primarily defined as positive or negative cases. It is critical to verify case-by-case by concentrating urine 10 fold by lyophilization for the diagnosis. Following lyophilization, persistent trace readings should be read as negatives. No trained technician is needed and cost is restricted to the cost of a lyophilizer and the electricity to run it.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Accuracy of Urine Circulating Cathodic Antigen Test for the Diagnosis of Schistosoma mansoni in Preschool-Aged Children before and after Treatment

BACKGROUND The Kato-Katz technique is widely used for the diagnosis of Schistosoma mansoni, but shows low sensitivity in light-intensity infections. We assessed the accuracy of a commercially available point-of-care circulating cathodic antigen (POC-CCA) cassette test for the diagnosis of S. mansoni in preschool-aged children before and after praziquantel administration. METHODOLOGY A 3-week ...

متن کامل

The use of circulating cathodic antigen rapid test and serology for diagnosis of active Schistosoma mansoni infection in migrants in Italy, a non-endemic country: a cross sectional study

Diagnosis of schistosomiasis in migrants coming from endemic areas can be difficult, especially in asymptomatic subjects. Light-intensity disease, in fact, may be missed due to the low sensitivity of the stool microscopy and serologic testing cannot distinguish between a resolved infection and an active infection in patients who have been infected and treated in the past, because specific antib...

متن کامل

Additional Evaluation of the Point-of-Contact Circulating Cathodic Antigen Assay for Schistosoma mansoni Infection

Studies of the urine-based point-of-contact cathodic circulating antigen test (POC-CCA) in Schistosoma mansoni-endemic settings in Africa indicate it has good sensitivity in detecting infections, but in areas of low prevalence, the POC-CCA can be positive for persons who are egg-negative by Kato-Katz stool assays. We examined the POC-CCA assay for: (a) batch-to-batch stability; (b) intra-reader...

متن کامل

Evaluation of a Urine Pooling Strategy for the Rapid and Cost-Efficient Prevalence Classification of Schistosomiasis

BACKGROUND A key epidemiologic feature of schistosomiasis is its focal distribution, which has important implications for the spatial targeting of preventive chemotherapy programs. We evaluated the diagnostic accuracy of a urine pooling strategy using a point-of-care circulating cathodic antigen (POC-CCA) cassette test for detection of Schistosoma mansoni, and employed simulation modeling to te...

متن کامل

Performance of POC-CCA® in diagnosis of schistosomiasis mansoni in individuals with low parasite burden.

INTRODUCTION Schistosomiasis, caused by Schistosoma mansoni, is a public health concern in Brazil. However, the most popular diagnostic method, the Kato-Katz technique, exhibits low sensitivity in low-endemicity areas. We aimed to compare the performance of an immunological assay, the point-of-care circulating cathodic antigen (POC-CCA®) test, with that of two parasitological techniques in a lo...