Interference by acetaminophen in the glucose oxidase-peroxidase method for blood glucose determination.

نویسندگان

  • I Kaufmann-Raab
  • H G Jonen
  • E Jähnchen
  • G F Kahl
  • U Groth
چکیده

Acetaminophen, p-aminophenol, and oxyphenbutazone interfere with the glucose oxidase/peroxidase method for glucose. Structurally related compounds that lack a free phenolic hydroxyl group (acetanilide, aniline, and phenylbutazone) do not interfere. During the analytical procedure acetaminophen is consumed. One mole of acetaminophen leads to an apparent loss of four moles of glucose. The hexokinase/glucose-6-phosphate dehydrogenase method (Boehringer Hexokinase method) is not affected by these substances.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Interference in an electrochemical detection system for peroxidase-linked reactions based on a fluoride ion-selective electrode.

Use of the fluoride ion-selective electrode as a detector in a linked glucose oxidase (EC 1.1.3.4)/peroxidase (EC 1.11.1.7) method for glucose was investigated for possible interference from sugars, metabolites, drugs, and anticoagulants. The CV for the method was 1.5% (SD 0.37 mmol/L) at a glucose concentration of 25.0 mmol/L. Interference was studied with glucose at this concentration, interf...

متن کامل

A Simple Image Analysis Method for Determination of Glucose by using Glucose Oxidase CdTe/TGA Quantum Dots

Glucose, as the major energy source in cellular metabolism, plays an important role in the natural growth of cells. Herein, a simple, rapid and low-cost method for the glucose determination by utilizing glucose oxidase and CdTe/thioglycolic acid (TGA) quantum dots (QDs) on a thin layer chromatography (TLC) plate has been described. The detection was based on the combination of the glucose enzym...

متن کامل

Modified glucose oxidase method for determination of glucose in whole blood.

We have modified an ultramicro method for determining glucose with glucose oxidase-peroxidase-odianisidine. Capillary blood is diluted in isotonic NaF (17.6-17.8 g/liter), the cells are removed, and glucose is determined after a 20-mm incubation with enzyme-containing reagent. Increasing the enzyme concentration fourfold nullifies interference from ascorbic acid and hemolysis, while greatly dec...

متن کامل

A New Method to Evaluate Fasting Plasma Glucose by Salivary Glucose Measurement

Objective:Fasting plasma glucose (FPG) analysis is the most importantmethod for detection and diagnosis of diabetes mellitus. Due to difficulty and problems of this method for determination of glycemia in diabetic patients, recently the use of Fasting Salivary Glucoseas a simple and non-invasive method to evaluate FPGhas came into significant considerationof specialists. The aim of this study w...

متن کامل

Methemalbumin interference in various biochemical assays of plasma.

During intravascular hemolysis and digestion of extravasated blood, heme from hemoglobin is oxidized to hematin, hich combines with plasma proteins to form methemalbu(1-4). One case of strong methemalbuminemia (140 oI/L) associated with intravascular hemolysis has been bserved in a patient after placing of an intracardiac valve. e plasma of this patient was so brown that we became terested in t...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical chemistry

دوره 22 10  شماره 

صفحات  -

تاریخ انتشار 1976