Relationship between catheter contact force and radiofrequency lesion size and incidence of steam pop in the beating canine heart: electrogram amplitude, impedance, and electrode temperature are poor predictors of electrode-tissue contact force and lesion size.

نویسندگان

  • Atsushi Ikeda
  • Hiroshi Nakagawa
  • Hendrik Lambert
  • Dipen C Shah
  • Edouard Fonck
  • Aude Yulzari
  • Tushar Sharma
  • Jan V Pitha
  • Ralph Lazzara
  • Warren M Jackman
چکیده

BACKGROUND Electrode-tissue contact force (CF) is believed to be a major factor in radiofrequency lesion size. The purpose of this study was to determine, in the beating canine heart, the relationship between CF and radiofrequency lesion size and the accuracy of predicting CF and lesion size by measuring electrogram amplitude, impedance, and electrode temperature. METHODS AND RESULTS Eight dogs were studied closed chest. Using a 7F catheter with a 3.5 mm irrigated electrode and CF sensor (TactiCath, St. Jude Medical), radiofrequency applications were delivered to 3 separate sites in the right ventricle (30 W, 60 seconds, 17 mL/min irrigation) and 3 sites in the left ventricle (40 W, 60 seconds, 30 mL/min irrigation) at (1) low CF (median 8 g); (2) moderate CF (median 21 g); and (3) high CF (median 60 g). Dogs were euthanized and lesion size was measured. At constant radiofrequency and time, lesion size increased significantly with increasing CF (P<0.01). The incidence of a steam pop increased with both increasing CF and higher power. Peak electrode temperature correlated poorly with lesion size. The decrease in impedance during the radiofrequency application correlated well with lesion size for lesions in the left ventricle but less well for lesions in the right ventricle. There was a poor relationship between CF and the amplitude of the bipolar or unipolar ventricular electrogram, unipolar injury current, and impedance. CONCLUSIONS Radiofrequencylesion size and the incidence of steam pop increase strikingly with increasing CF. Electrogram parameters and initial impedance are poor predictors of CF for radiofrequency ablation.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Locations of high contact force during left atrial mapping in atrial fibrillation patients: electrogram amplitude and impedance are poor predictors of electrode-tissue contact force for ablation of atrial fibrillation.

BACKGROUND During radiofrequency ablation, high electrode-tissue contact force (CF) is associated with increased risk of steam pop and perforation. The purpose of this study, in patients undergoing ablation of paroxysmal atrial fibrillation, was to: (1) identify factors producing high CF during left atrial (LA) and pulmonary vein mapping; (2) determine the ability of atrial potential amplitude ...

متن کامل

The Role Of Contact Force In Atrial Fibrillation Ablation.

During radiofrequency (RF) ablation, low electrode-tissue contact force (CF) is associated with ineffective RF lesion formation, whereas excessive CF may increase the risk of steam pop and perforation. Recently, ablation catheters using two technologies have been developed to measure real-time catheter-tissue CF. One catheter uses three optical fibers to measure microdeformation of a deformable...

متن کامل

Giant J-wave (Osborn wave) unrelated to hypothermia.

contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. Circ Arrhythm Electrophysiol 2008;1:354–62. 3. Shah DC, Leo G, Aeby N, Le Royer E, Sunthorn H, Burri H et al. Evaluation of a new catheter sensor for real-time measurement of tissue contact. Heart Rhythm 2006;3:S75–6. Abstract. 4. Ikeda A, Nakagawa H, Sha...

متن کامل

Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus.

BACKGROUND An open-irrigated radiofrequency (RF) ablation catheter was developed to measure contact force (CF). Three optical fibers measure microdeformation of the catheter tip. The purpose of this study was to (1) validate the accuracy of CF sensor (CFS) (bench test); and (2) determine the relationship between CF and tissue temperatures, lesion size, steam pop, and thrombus during RF ablation...

متن کامل

Target indices for clinical ablation in atrial fibrillation: insights from contact force, electrogram, and biophysical parameter analysis.

BACKGROUND In animal studies of radiofrequency ablation, lesion sizes plateau as the maximum lesion size is reached for an ablation. Lesion parameters are not available in clinical ablations, but preclinical work suggests that these correlate with impedance drop and electrogram attenuation. Characterization of the relationships between catheter contact force, ablation duration, and these surrog...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Circulation. Arrhythmia and electrophysiology

دوره 7 6  شماره 

صفحات  -

تاریخ انتشار 2014