Image‐guided bolus electron conformal therapy – a case study
نویسندگان
چکیده
We report on our initial experience with daily image guidance for the treatment of a patient with a basal cell carcinoma of the nasal dorsum using bolus electron conformal therapy. We describe our approach to daily alignment using treatment machine-integrated megavoltage (MV) planar imaging in conjunction with cone beam CT (CBCT) volumetric imaging to ensure the best possible setup reproducibility. Based on MV imaging, beam aperture misalignment with the intended treatment region was as large as 0.5 cm in the coronal plane. Four of the five fractions analyzed show induced shifts when compared to digitally reconstructed radiographs (DRR), in the range of 0.2-0.5 cm. Daily inspection of CBCT images show that the bolus device can have significant tilt in any given direction by as much as 13° with respect to beam axis. In addition, we show that CBCT images reveal air gaps between bolus and skin that vary from day to day, and can potentially degrade surface dose coverage. Retrospective dose calculation on CBCT image sets shows that when daily shifts based on MV imaging are not corrected, geometrical miss of the planning target volume (PTV) can cause an underdosing as large as 14% based on DVH analysis of the dose to the 90% of the PTV volume.
منابع مشابه
On Presentation of Optimal Treatment Plan in Radiotherapy of Parotid Cancer: A Comparison of Nine Techniques in Three Dimensional Conformal Radiation Therapy
Introduction: Today, radiotherapy combined with surgery in the treatment of parotid tumors widely are used specially for high grade parotid tumors. One of concern and challenge issue of radiotherapy is that the treatment of these tumors with irregular surface due to the presence of external ear and region of different physical electron density (air cavities, dense bone, soft t...
متن کاملEffect of Bolus Frequency and Its Thickness in Postmastectomy Three-dimensional Conformal Radiotherapy on Skin Dose for Superposition Algorithm
Introduction: The postmastectomy radiotherapy uses bolus to improve the coverage close to the skin; however, it needs to be removed in case of severe skin toxicity. This study investigated the effect of bolus parameters (i.e., frequency and thickness) for the superposition algorithm on skin dose in postmastectomy three-dimensional conformal radiotherapy (3D-CRT). <str...
متن کاملComparison of three different techniques in parotid gland tumors using three dimensional conformal radiation therapy
Introduction: Background: Radiotherapy of parotid tumors is achieved by different radiotherapy techniques. So they deliver the dose to organs at risk such as contralateral parotid, temporal lobe & spinal cord. The consequences of high dose to OARs are xerostomia, Temporal lobe necrosis and loss of hearing. Purpose: The aim of this study was to compare different treatment tech...
متن کاملHealth technology assessment of image-guided radiotherapy (IGRT): A systematic review of current evidence
Background: Image-guided radiotherapy used multiple imaging during the radiation therapy course to improve the precision and accuracy of health care provider's treatment. Objectives: This study aims to assess the safety, effectiveness and economic aspects of image-guided radiation therapy for decision-making about this technology in Iran. Methods: In this study, the most important med...
متن کاملنقش تکنولوژی در بهبود درمان سرطان
One major strategy, technology-driven improvement of treatment conformity in cancer treatment, including advanced image guidance, advanced charged particle therapy (CPT), and application of nanoparticles in hyperthermia, will enable further widening of the therapeutic window of cancer treatment in the era of precision medicine. The state of the art treatment in photon-therapy is advanced 3D...
متن کامل