Flexibility and trackability of laser cut coronary stent systems.

نویسندگان

  • Péter Szabadíts
  • Zsolt Puskás
  • János Dobránszky
چکیده

Coronary stents are the most important supports in present day cardiology. Flexibility and trackability are two basic features of stents. In this paper, four different balloon-expandable coronary stent systems were investigated mechanically in order to compare their suitability. The coronary stent systems were assessed by measurements of stent flexibility as well as by comparison of forces during simulated stenting in a self-investigated coronary vessel model. The stents were cut by laser from a single tube of 316L stainless steel or L-605 (CoCr) cobalt chromium alloy. The one-and four-point bending tests were carried out to evaluate the stent flexibility E x I (Nmm(2)), under displacement control in crimped and expanded configurations. The flexibility of stents would be rather dependent on the design than on raw material. In general a more flexible stent needs lower tracking force during the implantation. The L-605 raw material stents need lower track force to pass through in the vessel model than the 316L raw material stents. The sort and long stents passed through the curved vessel model in different ways. The long stents nestled to the vessel wall at the outer arc and bent, while the short stents did not bend in the curve, only the delivery systems bent.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Development of gold stents for the treatment of intracranial aneurysms: an experimental study in a canine model.

BACKGROUND AND PURPOSE Gold has often been used in medicine because of its radiopacity and flexibility. To perform stent-supported coil embolization of intracranial aneurysms, we prepared a gold stent and examined its flexibility, radiopacity, and thrombogenic properties in comparison with a stainless steel device implanted in vitro and in vivo. METHODS Gold stents were prepared by plating go...

متن کامل

In vitro and human studies of a 4F double-coaxial technique ("mother-child" configuration) to facilitate stent implantation in resistant coronary vessels.

BACKGROUND We recently developed a 4F child catheter that can be inserted into 6F or larger conventional guiding catheters. The use of 4F mother-child technique may improve the delivery of coronary stents to complex lesions. Accordingly, we sought to determine the potential of using a 4F mother-child technique to treat complex coronary lesions. METHODS AND RESULTS The support power and the tr...

متن کامل

Emergency coronary angioplasty with stenting using Cordis® diagnostic coronary catheters when there is difficulty in engaging guide catheters and bench evaluation of diagnostic and guide catheters.

INTRODUCTION AND AIMS Difficulty in engaging with guide catheters is not uncommon in acute emergencies. We aimed to evaluate the use of Cordis® INFINITI diagnostic catheters to perform angioplasty in patients in whom the coronaries cannot be engaged using standard guide catheters. METHODS In 34 cases of acute coronary syndrome, when difficulty in engagement with two standard guide catheters w...

متن کامل

Management of a large coronary artery aneurysm with graft-coated stent during the acute phase of myocardial infarction

Coronary aneurysms in adults are rare clinical entities. Herein, we presented a 54-year-old man who was admitted with an acute extensive anterior myocardial infarction. The emergency coronary angiography revealed an isolated large aneurysm at the proximal segment of the left anterior descending coronary artery. The patient was successfully treated with a graft-coated stent

متن کامل

Association of G22A variant of Adenosine Deaminase gene with coronary in-stent restenosis in coronary artery patients receiving drug-eluting stent

In-stent restenosis (ISR) is regarded as the main problem in the utilization of stents in the treatment of coronary artery atherosclerotic stenosis in percutaneous coronary intervention (PCI). This study investigated the possible role of the G22A variant of the Adenosine Deaminase gene (ADA) in the development of ISR. In this study, 91 patients who underwent PCI were divided into two groups of ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Acta of bioengineering and biomechanics

دوره 11 3  شماره 

صفحات  -

تاریخ انتشار 2009