Increase in the Tibial Slope Reduces Wear after Medial Unicompartmental Fixed-Bearing Arthroplasty of the Knee

نویسندگان

  • Patrick Weber
  • Christian Schröder
  • Jens Schwiesau
  • Sandra Utzschneider
  • Arnd Steinbrück
  • Matthias F Pietschmann
  • Volkmar Jansson
  • Peter E Müller
چکیده

Introduction. Unicompartmental arthroplasty of the knee in patients with isolated medial osteoarthritis gives good results, but survival is inferior to that of total knee prosthesis. Knees may fail because positioning of the prosthesis has been suboptimal. The aim of this study was to investigate the influence of the tibial slope on the rate of wear of a medial fixed-bearing unicompartmental knee arthroplasty. Materials and Methods. We simulated wear on a medial fixed-bearing unicompartmental knee prosthesis (Univation) in vitro with a customised, four-station, and servohydraulic knee wear simulator, which exactly reproduced the walking cycle (International Organisation for Standardisation (ISO) 14243-1:2002(E)). The medial prostheses were inserted with 3 different posterior tibial slopes: 0°, 4°, and 8° (n = 3 in each group). Results. The wear rate decreased significantly between 0° and 4° slope from 10.4 (SD 0.62) mg/million cycles to 3.22 (SD 1.71) mg/million cycles. Increasing the tibial slope to 8° did not significantly change the wear rate. Discussion. As an increase in the tibial slope reduced the wear rate in a fixed-bearing prosthesis, a higher tibial slope should be recommended. However, other factors that are influenced by the tibial slope (e.g., the tension of the ligament) must also be considered.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of implant position and knee alignment after patient-specific unicompartmental knee arthroplasty.

Implant positioning and knee alignment are two primary goals of successful unicompartmental knee arthroplasty. This prospective study outlines the radiographic results following 32 patient-specific unicompartmental medial resurfacing knee arthroplasties. By means of standardized pre- and postoperative radiographs of the knee in strictly AP and lateral view, AP weight bearing long leg images as ...

متن کامل

Unicompartmental Knee Osteoarthritis (UKOA): Unicompartmental Knee Arthroplasty (UKA) or High Tibial Osteotomy (HTO)?

  Background: The aim of this review article is to analyze the results of high tibial osteotomy compared to unicompartmental knee arthroplasty in patients with unicompartmental knee osteoarthritis.   Methods: The search engine used was PubMed. The keywords were: "high tibial osteotomy versus unicompartmental knee arthroplasty". Twenty-one articles were found on 28 February 2015, but only eighte...

متن کامل

In-vitro wear of fixed and mobile bearing unicompartmental knee replacements

Introduction Unicompartmental knee replacement (UKR) is an increasingly popular surgical intervention for the treatment of osteoarthritis, involving the replacement of only one compartment of the natural knee. There is promising clinical literature which shows excellent long term survivorship of both fixed and mobile bearing UKR, with fixed bearings showing more consistent long-term performance...

متن کامل

Early Results of Oxford Mobile Bearing Medial Unicompartmental Knee Replacement (UKR) with the Microplasty Instrumentation: An Indian Experience

Background: Oxford medial unicompartmental knee replacement (UKR) is indicated in patients with anteromedialosteoarthritis (AMOA) of the knee. Microplasty (MP) instrumentation was introduced in 2012 as an improvement overphase 3 instrumentation. Advantages of this instrumentation include conservative tibial cut, decreased tibial re-cut rateand improved component alignment. We report the results...

متن کامل

Effect of Polyethylene Crosslinking and Bearing Design on Wear of Unicompartmental Arthroplasty.

Wear and polyethylene damage continue to be important factors affecting outcomes of unicompartmental knee arthroplasty. We compared two design rationales for unicompartmental arthroplasty: fully congruent mobile bearings; or moderately conforming fixed bearings using experimental and computational wear simulation. Experimental wear rates were 3.89 (±0.12) mg/million cycles for the highly crossl...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 2015  شماره 

صفحات  -

تاریخ انتشار 2015