Fabrication of a Micro-Needle Array Electrode by Thermal Drawing for Bio-Signals Monitoring
نویسندگان
چکیده
A novel micro-needle array electrode (MAE) fabricated by thermal drawing and coated with Ti/Au film was proposed for bio-signals monitoring. A simple and effective setup was employed to form glassy-state poly (lactic-co-glycolic acid) (PLGA) into a micro-needle array (MA) by the thermal drawing method. The MA was composed of 6 × 6 micro-needles with an average height of about 500 μm. Electrode-skin interface impedance (EII) was recorded as the insertion force was applied on the MAE. The insertion process of the MAE was also simulated by the finite element method. Results showed that MAE could insert into skin with a relatively low compression force and maintain stable contact impedance between the MAE and skin. Bio-signals, including electromyography (EMG), electrocardiography (ECG), and electroencephalograph (EEG) were also collected. Test results showed that the MAE could record EMG, ECG, and EEG signals with good fidelity in shape and amplitude in comparison with the commercial Ag/AgCl electrodes, which proves that MAE is an alternative electrode for bio-signals monitoring.
منابع مشابه
Fabrication of Micro-Needle Electrodes for Bio-Signal Recording by a Magnetization-Induced Self-Assembly Method
Micro-needle electrodes (MEs) have attracted more and more attention for monitoring physiological electrical signals, including electrode-skin interface impedance (EII), electromyography (EMG) and electrocardiography (ECG) recording. A magnetization-induced self-assembling method (MSM) was developed to fabricate a microneedle array (MA). A MA coated with Ti/Au film was assembled as a ME. The fr...
متن کاملFabrication of Polymeric Micro needle Array by Micromachining & Micro molding
Microneedles are expected to have greater demand in biomedical applications like sensing, electroporation & drug delivery in the future and there is need for development of fabrication process which is simple, economical and productive. Present work discusses about fabrication of PDMS (polydimethylsiloxane) micro needle array electrode for bio-potential signal measurement like EEG by micromachi...
متن کاملDesign and Simulation of a Fluidic Micro-Bio-Sensor Based on Resonator Array
In this paper, a fluidic biosensor with possibility to fabricate by Micro-Electro-Mechanical Systems (MEMS) technology is proposed for biomedical mass detection and lab-on-chip applications. This is designed by electromechanical coupling of harmonic micromechanical resonators with harmonic springers as a mechanical resonator array. It can disperse mechanical wave along the array by electrostati...
متن کاملFlexible Electrode Array for Retinal Stimulation
In this Work, ITO/PET (Indium Tin Oxide / Polyethylene Terephthalate) electrode structure which provides biocompatibility, mechanical stability and flexibility is fabricated. Flexible ITO/PET implantable electrode array for a retina has been developed. The electrode array is fabricated on a thin PET/ITO substrate and is encapsulated using, SU-8, an insulating material. PET substrate and SU-8 po...
متن کاملA Short Review on Fabrication Methods of Micro-Cantilever for Ionic Electroactive Polymer Sensors/Actuators
Micro-electroactive polymers have wide applications from industry to healthcare. Artificial muscles, microvalves, microswitches, haptic sensing, blood pressure and pulse monitoring are some potential applications for micro-electroactive polymers. The most advantage of micro-electroactive polymers as sensors and actuators is bio-compatibility and low power threshold for large deformations...
متن کامل