Activation and entrainment mapping of hemodynamically unstable ventricular tachycardia using a percutaneous left ventricular assist device.
نویسندگان
چکیده
OBJECTIVES Our goal was to investigate the effects of percutaneous left ventricular assist device (pLVAD) support during catheter ablation of unstable ventricular tachycardia (VT). BACKGROUND Mechanical cardiac support during ablation of unstable VT is being increasingly used, but there is little available information on the potential hemodynamic benefits. METHODS Twenty-three consecutive procedures in 22 patients (ischemic, n = 11) with structural heart disease and hemodynamically unstable VT were performed with either pLVAD support (n = 10) or no pLVAD support (intra-aortic balloon pump counterpulsation, n = 6; no support, n = 7). Procedural monitoring included vital signs, left atrial pressure, arterial blood pressure, cerebral perfusion/oximetry, VT characteristics, and ablation outcomes. RESULTS The pLVAD group was maintained in VT significantly longer than the non-pLVAD group (66.7 min vs. 27.5 min; p = 0.03) and required fewer early terminations of sustained VT for hemodynamic instability (1.0 vs. 4.0; p = 0.001). More patients in the pLVAD group had at least 1 VT termination during ablation than non-pLVAD patients (9 of 10 [90%] vs. 5 of 13 [38%]; p = 0.03). There were no differences between groups in duration of cerebral deoxygenation, hypotension or perioperative changes in left atrial pressure, brain natriuretic peptide levels, lactic acid, or renal function. CONCLUSIONS In patients with scar-related VT undergoing catheter ablation, pLVAD support was able to safely maintain end-organ perfusion despite extended periods of hemodynamically unstable VT. Randomized studies are necessary to determine whether this enhanced ability to perform entrainment and activation mapping will translate into a higher rate of clinical success.
منابع مشابه
Percutaneous left ventricular assist devices in ventricular tachycardia ablation: multicenter experience.
BACKGROUND Data on relative safety, efficacy, and role of different percutaneous left ventricular assist devices for hemodynamic support during the ventricular tachycardia (VT) ablation procedure are limited. METHODS AND RESULTS We performed a multicenter, observational study from a prospective registry including all consecutive patients (N=66) undergoing VT ablation with a percutaneous left ...
متن کاملThe role of percutaneous left ventricular assist devices during ventricular tachycardia ablation.
Ventricular tachycardia (VT) is a common but serious arrhythmia that significantly adds to the morbidity and mortality of patients with structural heart disease. Percutaneous catheter ablation has evolved to be standard therapy to prevent recurrent implantable cardioverter defibrillator shocks from VT in patients on antiarrhythmia medications. Procedural outcomes in patients with structural hea...
متن کاملIntramural Reentrant Ventricular Tachycardia in a Patient with Severe Hypertensive Left Ventricular Hypertrophy
We describe the case of a patient with severe hypertensive left ventricular hypertrophy and sustained hemodynamically unstable ventricular tachycardia (VT). Entrainment was demonstrated in the electrophysiological study. Activation mapping and pacemapping identified the location of the intramural reentrant VT with the exit site close to the epicardium. However, VT persisted after ablation at th...
متن کاملCatheter ablation of hemodynamically unstable ventricular tachycardia in ischemic cardiomyopathy using high‐resolution mapping
Catheter ablation is a recommended therapy option for ventricular tachycardia (VT). The antegrade transseptal approach for targeting VT with left ventricular origin is feasible with the high-resolution basket catheter. High-resolution mapping offers the potential to quickly acquire detailed voltage and activation maps. This may help to identify the crucial VT-substrate even in patients with hug...
متن کاملThe Role of Intravenous Dopamine on Hemodynamic Support during Radiofrequency Catheter Ablation of Poorly Tolerated Idiopathic Ventricular Tachycardia
BACKGROUND AND OBJECTIVES Hemodynamically unstable idiopathic ventricular tachycardias (VTs) are a challenge for activation or entrainment mapping technique. Mechanical circulatory support is an option, but is not always readily available. In this study, we investigated the safety and efficacy of hemodynamic support using intravenous (IV) dopamine solely during radiofrequency catheter ablation ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of the American College of Cardiology
دوره 58 13 شماره
صفحات -