Tono-Pen tonometry in normal and in post-keratoplasty eyes.

نویسندگان

  • O Geyer
  • Y Mayron
  • A Loewenstein
  • M Neudorfer
  • L Rothkoff
  • M Lazar
چکیده

Oculab Tono-Pen tonometry was compared with Goldmann applanation tonometry in 82 eyes of 82 patients with normal corneas and in 54 eyes of 54 patients who had undergone penetrating keratoplasty and whose corneas did not preclude the use of Goldmann tonometer. We found that the intraocular pressure (IOP) in 48% of the eyes with normal corneas and in 57% after keratoplasty has different measurements with Goldmann and Tono-Pen pressures of 3 mm Hg or more. Despite the correlation between the Goldmann tonometer and the Tono-Pen in the group of eyes with normal corneas (r = 0.83) as well as in the group of eyes after keratoplasty (r = 0.79) the Tono-Pen tended to significantly overestimate the Goldmann tonometer reading (p < 0.0001). The mean difference between the two instruments was highest across the lower IOP range (< 9 mm Hg) in the group of eyes after keratoplasty. Because the mean absolute values of the paired differences between Goldmann and Tono-Pen measurements varied significantly across all IOP intervals it was not possible to establish a correction factor which could be used when comparing the two measurements. Based on this study the Tono-Pen consistently overestimated the actual IOP in an unpredictable manner. Where possible Goldmann measurements of the IOP are still to be preferred.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Intraocular pressure measurement after penetrating keratoplasty: minified Goldmann applanation tonometer, pneumatonometer, and Tono-Pen versus manometry.

The accuracy of intraocular pressure measurement with the minified Goldmann applanation tonometer, the pneumatonometer, and the Tono-Pen tonometer were compared in post-mortem human eyes which had undergone penetrating keratoplasty. Enucleated post-mortem human eyes underwent same sized (7.75 mm) or 0.5 mm oversized (8.25 mm) autologous penetrating keratoplasty. Intraocular pressure was then se...

متن کامل

Tono-Pen determination of intraocular pressure in patients with band keratopathy or glued cornea.

AIMS To evaluate the intraocular pressure (IOP) measurements in patients with band keratopathy or glued corneas obtained from affected and non-affected areas. METHODS 15 patients with band keratopathy or glued corneas were prospectively recruited. When both eyes were affected, only the right eye was analysed. Tono-Pen readings of IOP were obtained sequentially from the affected and non-affect...

متن کامل

Clinical comparison of the ProTon and Tono-Pen tonometers with the Goldmann applanation tonometer.

A clinical evaluation of a new electron ProTon tonometer was performed comparing the values of intraocular pressure (IOP) measured using this instrument with those determined by a similar instrument, Tono-Pen XL, and by Goldmann applanation tonometry. The mean IOP measured in 106 eyes with the ProTon tonometer was not significantly different from that determined with Goldmann applanation, while...

متن کامل

Tono-Pen XL tonometry during application of a suction ring in rabbits

BACKGROUND The purpose of this study is to evaluate the use of Tono-Pen XL in measuring IOP during the application of a suction ring in rabbit eyes with manometrically controlled IOP. METHODS Tono-Pen XL was calibrated against direct manometry in 10 rabbit eyes. A suction ring was then applied in 4 rabbit eyes and the IOP was determined manometrically during suction ring application at 350 mm...

متن کامل

Validity of Tono-pachymetry for Measuring Corrected Intraocular Pressure in Non-surgical and Post-photorefractive Keratectomy Eyes

PURPOSE To assess the validity of central corneal thickness (CCT) and corrected intraocular pressure (IOP) values obtained by tono-pachymetry in non-surgical and post-photorefractive keratectomy (PRK) eyes. METHODS For the study, 108 young healthy participants and 108 patients who had PRK were enrolled. Measurements were randomly performed by tono-pachymetry, ultrasonic (US) pachymetry, and G...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The British journal of ophthalmology

دوره 76 9  شماره 

صفحات  -

تاریخ انتشار 1992