Oxygen delivery using a neonatal self-inflating resuscitation bag: effect of oxygen flow.
نویسندگان
چکیده
OBJECTIVE We evaluated the effect of oxygen (O₂) flow rate on the corresponding delivered fraction of oxygen (FiO₂) during positive pressure ventilation (PPV) when using a neonatal self-inflating bag (SIB). METHODS Fifteen health care professionals administered PPV at a respiratory rate of 40 to 60 breaths per minute and at peak inspiratory pressures of 25 and 35 cm H₂O to a manikin by using a SIB with reservoir connected to an O₂ source equipped with a flowmeter (flow rates: 0-10 L/min). The FiO₂ corresponding to each flow rate was measured at the inflow to the facial mask for 60 seconds. RESULTS In total, 2520 FiO₂ data points were collected. At every O₂ flow rate, the FiO₂ gradually increased from time 0 seconds to time 60 seconds, both at 25 cm H₂O and at 35 cm H₂O. After 1 minute of PPV at 25 cm H₂O, the delivered FiO₂ was 31.5% ± 2.1% and 43.1% ± 3.1% at O₂ flow rates of 0.1 and 0.5 L/min, respectively. After 1 minute of PPV at 35 cm H₂O, the delivered FiO₂ was 29.4% ± 2.0% and 42.1% ± 4.6% at O₂ flow rates of 0.1 and 0.5 L/min, respectively. At all O₂ flow rates >5 L/min, the delivered FiO₂ was >85% and >95%, after 1 minute of PPV at 25 and 35 cm H₂O, respectively. CONCLUSIONS Delivered FiO₂ during PPV depends on 3 factors: oxygen flow rate, peak inspiratory pressures, and time elapsed. These data can be used to develop a scheme correlating the oxygen flow rate and the corresponding delivered FiO₂ when using a neonatal SIB.
منابع مشابه
The self-inflating resuscitation bag delivers high oxygen concentrations when used without a reservoir: implications for neonatal resuscitation.
OBJECTIVE To measure the delivered fractional oxygen concentration (F(DO(2))) from preterm-size Laerdal silicone resuscitators (PLSR) without a reservoir. BACKGROUND The North American Neonatal Resuscitation Program manual states that self-inflating bags without a reservoir deliver approximately 40% oxygen, differing from the PLSR manufacturer's specifications. METHODS A neonatal test lung ...
متن کاملOxygen outflow delivered by manually operated self-inflating resuscitation bags in patients breathing spontaneously.
OBJECTIVE To determine the oxygen outflow delivered by seven different models of manually operated self-inflating resuscitation bags (with and without an oxygen reservoir connected), which were tested using different oxygen supply rates without manipulating the bag, by simulating their use in patients breathing spontaneously. METHODS The oxygen outflow was measured using a wall oxygen flow me...
متن کاملA survey of delivery room resuscitation practices in the United States.
OBJECTIVE To determine current resuscitation practices of neonatologists in the United States. METHODS A 15-question survey was developed and mailed to neonatal directors in May 2004. RESULTS Of the total of 797 surveys mailed, 84 were returned undeliverable or unanswered and 450 were returned completed (63% response rate). Respondents were mainly (70%) from level III NICUs. Most programs r...
متن کاملNeonatal and pediatric manual hyperinflation: influence of oxygen flow on ventilation parameters.
BACKGROUND Although self-inflating bags are widely used for manual hyperinflation, they do not allow ventilation parameters, such as pressure or volume, to be set. We studied the ventilation performance of neonatal and pediatric self-inflating bags. METHODS We asked 22 physiotherapists to manually hyperinflate 2 lung models (neonatal and pediatric), using self-inflating bags from 3 manufactur...
متن کاملAlterations in peak inspiratory pressure and tidal volume delivered by manually operated self-inflating resuscitation bags as a function of the oxygen supply rate.
OBJECTIVE To assess possible alterations in the tidal volume and peak inspiratory pressure delivered by seven models of manually operated self-inflating resuscitation bags as a function of the oxygen supply rate used. METHODS The resuscitation bags tested were the following: Oxigel, models A and B; CE Reanimadores; ProtecSolutions; Missouri; Axmed; and Narcosul. For the measurements, a wall o...