Comparison of standard cardiopulmonary resuscitation versus the combination of active compression-decompression cardiopulmonary resuscitation and an inspiratory impedance threshold device for out-of-hospital cardiac arrest.

نویسندگان

  • Benno B Wolcke
  • Dietmar K Mauer
  • Mark F Schoefmann
  • Heinke Teichmann
  • Terry A Provo
  • Karl H Lindner
  • Wolfgang F Dick
  • Dorothee Aeppli
  • Keith G Lurie
چکیده

BACKGROUND Active compression-decompression (ACD) CPR combined with an inspiratory impedance threshold device (ITD) improves vital organ blood flow during cardiac arrest. This study compared survival rates with ACD+ITD CPR versus standard manual CPR (S-CPR). METHODS AND RESULTS A prospective, controlled trial was performed in Mainz, Germany, in which a 2-tiered emergency response included early defibrillation. Patients with out-of-hospital arrest of presumed cardiac pathogenesis were sequentially randomized to ACD+ITD CPR or S-CPR by the advanced life support team after intubation. Rescuers learned which method of CPR to use at the start of each work shift. The primary end point was 1-hour survival after a witnessed arrest. With ACD+ITD CPR (n=103), return of spontaneous circulation and 1- and 24-hour survival rates were 55%, 51%, and 37% versus 37%, 32%, and 22% with S-CPR (n=107) (P=0.016, 0.006, and 0.033, respectively). One- and 24-hour survival rates in witnessed arrests were 55% and 41% with ACD+ITD CPR versus 33% and 23% in control subjects (P=0.011 and 0.019), respectively. One- and 24-hour survival rates in patients with a witnessed arrest in ventricular fibrillation were 68% and 58% after ACD+ITD CPR versus 27% and 23% after S-CPR (P=0.002 and 0.009), respectively. Patients randomized > or =10 minutes after the call for help to the ACD+ITD CPR had a 3 times higher 1-hour survival rate than control subjects (P=0.002). Hospital discharge rates were 18% after ACD+ITD CPR versus 13% in control subjects (P=0.41). In witnessed arrests, overall neurological function trended higher with ACD+ITD CPR versus control subjects (P=0.07). CONCLUSIONS Compared with S-CPR, ACD+ITD CPR significantly improved short-term survival rates for patients with out-of-hospital cardiac arrest. Additional studies are needed to evaluate potential long-term benefits of ACD+ITD CPR.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of an impedance threshold device in patients receiving active compression-decompression cardiopulmonary resuscitation for out of hospital cardiac arrest.

AIMS The purpose of this multicentre clinical randomized controlled blinded prospective trial was to determine whether an inspiratory impedance threshold device (ITD), when used in combination with active compression-decompression (ACD) cardiopulmonary resuscitation (CPR), would improve survival rates in patients with out-of-hospital cardiac arrest. METHODS AND RESULTS Patients were randomize...

متن کامل

Active compression-decompression resuscitation and impedance threshold device for out-of-hospital cardiac arrest: a systematic review and metaanalysis of randomized controlled trials.

OBJECTIVE Active compression-decompression resuscitation and impedance threshold device are proposed to improve survival of patients of cardiac arrest by lowering intrathoracic pressure and increasing cardiac output. The results of clinical studies of active compression-decompression resuscitation or impedance threshold device were controversial. This metaanalysis pooled results of randomized c...

متن کامل

Use of an inspiratory impedance threshold device on a facemask and endotracheal tube to reduce intrathoracic pressures during the decompression phase of active compression-decompression cardiopulmonary resuscitation.

INTRODUCTION Use of an inspiratory impedance threshold device (ITD) significantly increases coronary perfusion pressures and survival in patients ventilated with an endotracheal tube (ETT) during active compression-decompression cardiopulmonary resuscitation. We tested the hypothesis that the ITD could lower intratracheal pressures when attached to either a facemask or ETT. METHODS An active ...

متن کامل

Use of an inspiratory impedance threshold valve during cardiopulmonary resuscitation: a progress report.

Building upon studies on the mechanism of active compression-decompression (ACD) cardiopulmonary resuscitation, a new inspiratory impedance threshold valve has been developed to enhance the return of blood to the thorax during the decompression phase of CPR. Use of this device results in a greater negative intrathoracic pressure during chest wall decompression. This leads to improved vital orga...

متن کامل

Clinical Investigations Clinical evaluation of an inspiratory impedance threshold device during standard cardiopulmonary resuscitation in patients with out-of-hospital cardiac arrest*

An inspiratory impedance threshold device (ITD) has recently been demonstrated to enhance vital organ perfusion and neurologically intact survival rates in animals in cardiac arrest during standard cardiopulmonary resuscitation (CPR) (1– 4). The ITD is a small, 35-mL device that can attach to a variety of airway adjuncts, including a facemask and endotracheal tube (Fig. 1). By selectively imped...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Circulation

دوره 108 18  شماره 

صفحات  -

تاریخ انتشار 2003