Validation of MTF measurement for digital mammography quality control.
نویسندگان
چکیده
The modulation transfer function (MTF) describes the spatial resolution properties of imaging systems. In this work, the accuracy of our implementation of the edge method for calculating the presampled MTF was examined. Synthetic edge images with known MTF were used as gold standards for determining the robustness of the edge method. These images simulated realistic data from clinical digital mammography systems, and contained intrinsic system factors that could affect the MTF accuracy, such as noise, scatter, and flat-field nonuniformities. Our algorithm is not influenced by detector dose variations for MTF accuracy up to 1/2 the sampling frequency. We investigated several methods for noise reduction, including truncating the supersampled line spread function (LSF), windowing the LSF, applying a local exponential fit to the LSF, and applying a monotonic constraint to the supersampled edge spread function. Only the monotonic constraint did not introduce a systematic error; the other methods could result in MTF underestimation. Overall, our edge method consistently computed MTFs which were in good agreement with the true MTF. The edge method was then applied to images from a commercial storage-phosphor based digital mammography system. The calculated MTF was affected by the size (sides of 2.5, 5, or 10 cm) and the composition (lead or tungsten) of the edge device. However, the effects on the MTF were observed only with regard to the low frequency drop (LFD). Scatter nonuniformity was dependent on edge size, and could lead to slight underestimation of LFD. Nevertheless, this negative effect could be minimized by using an edge of 5 cm or larger. An edge composed of lead is susceptible to L-fluorescence, which causes overestimation of the LFD. The results of this work are intended to underline the need for clear guidelines if the MTF is to be given a more crucial role in acceptance tests and routine assessment of digital mammography systems: the MTF algorithm and edge object test tool need to be publicly validated.
منابع مشابه
محاسبۀ تابع انتقال مدولاسیون برای بررسی کنترل کیفی در رادیولوژی دیجیتال
The usage of digital radiology systems with insufficient quality control of digital image detectors can efface inherent superiorities of digital systems in comparison with analogue radiology systems and even makes an increased level of patients’ radiation doses. Modulation transfer function (MTF) is one of the most important parameters related to image quality and digital detectors efficiency. ...
متن کاملBreast dose in conventional and digital mamography
Introduction: The amount of doses received in mammography is one of the biggest concerns in the world. According to some statistics, every 12 minutes a woman dies due to breast cancer. Mammography concerned to detect breast cancer at an early stage of the disease as well as for screening, however due to the risk of exposure, the increasingly and widespread use of mammography fo...
متن کامل[Evaluation of the 1Shot Phantom dedicated to the mammography system using FCR].
Currently daily quality control (QC) tests for mammography systems are generally evaluated by using visual analysis phantoms, which of course means subjective measurement. In our study, however, we evaluated a novel digital phantom, the 1Shot Phantom M plus (1Shot Phantom), together with automatic analysis software dedicated for mammography systems using Fuji computed radiography (FCR). The dig...
متن کاملDigital Mammography: From Physical Performance Evaluation To Image Quality Analysis
Since the early 1990s digital X-ray detectors have been used extensively for medical imaging applications, such as Mammography, General Radiography, Computed Tomography, Tomosynthesis, Fluoroscopy, etc. The quality of a radiograph needs to be adequate to provide the required information for a given task. The primary physical parameters that affect image quality are spatial resolution, noise, an...
متن کاملمقایسه میزان دز جذبی پستان در دستگاه ماموگرافی معمولی و دیجیتال در شرایط مورد استفاده دستگاهها در تبریز
Background: The amount of doses received in mammography is one of the biggest concerns in the world. According to some statistics, every 12 minutes a woman dies due to breast cancer. Mammography concerned to detect breast cancer at an early stage of the disease as well as for screening, however due to the risk of exposure, the increasingly and widespread use of mammography for breast cancer scr...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Medical physics
دوره 32 6 شماره
صفحات -
تاریخ انتشار 2005