Calibration of the HemoCue point-of-care analyser for determining haemoglobin concentration in a lizard and a fish
نویسندگان
چکیده
Haemoglobin concentration ([Hb]) is measured for a wide variety of animal studies. The use of point-of-care devices, such as the HemoCue, is becoming increasingly common because of their portability, relative ease of use and low cost. In this study, we aimed to determine whether the [Hb] of blue-tongued skink (Tiliqua nigrolutea) blood can be determined accurately using the HemoCue and whether the HemoCue overestimates the [Hb] of reptile blood in a similar manner to fish blood. Additionally, we aimed to test whether ploidy affected [Hb] determined by the HemoCue using blood from diploid and triploid Atlantic salmon (Salmo salar). The HemoCue Hb 201(+) systematically overestimated [Hb] in both blue-tongued skinks and Atlantic salmon, and there was no difference between calibration equations determined for diploid or triploid salmon. The overestimation was systematic in both species and, as such, [Hb] determined by the HemoCue can be corrected using appropriate calibration equations.
منابع مشابه
Comparison of Point of Care (POC) Testing of Glucose by B Braun Glucometer and Hemocue Glucose 201+ Analyser Versus Centralised Testing in Neonatal Intensive Care Unit (NICU).
BACKGROUND Neonatal hypoglycemia is the most common carbohydrate metabolic disturbance seen in case of neonates and especially in preterm neonates. Accurate and rapid determination of hypoglycemia and its prompt treatment is of utmost importance to decrease morbidity and mortality of neonates. AIMS To estimate blood glucose in neonates and test the efficacy of HemoCue Glucose 201+ analyser an...
متن کاملEvaluation of the Haemoglobin Colour Scale and comparison with the HemoCue haemoglobin assay.
OBJECTIVE To evaluate the Haemoglobin Colour Scale developed by WHO for estimating haemoglobin concentration and to compare the results obtained using it and the HemoCue assay with those determined using a reference method, the Technicon H3 analyser. METHODS The Colour Scale and HemoCue assay were used to test 408 blood samples. Subsequently, Bland-Altman plots were determined and the proximi...
متن کاملEvaluation of the validity of a rapid method for measuring high and low haemoglobin levels in whole blood donors.
BACKGROUND Haemoglobin screening methods need to be highly sensitive to detect both low and high haemoglobin levels and avoid unnecessary rejection of potential blood donors. The aim of this study was to evaluate the accuracy of measurements by HemoCue in blood donors. MATERIALS AND METHODS Three hundred and fourteen randomly selected, prospective blood donors were studied. Single fingerstick...
متن کاملCost-effective on-site screening for anaemia in pregnancy in primary care clinics.
OBJECTIVE To determine the feasibility, accuracy and cost of developing a system of on-site screening for anaemia in pregnancy in primary care clinics. SETTING Mobile clinic team in Hlabisa health district, KwaZulu-Natal. METHODS Four hundred and forty-nine consecutive women attending antenatal clinics were screened for anaemia (haemoglobin < 10 g/dl) using copper sulphate solution; the res...
متن کاملDetermining the Lethal Concentration of Formalin to Koi fish (Cyprinus carpio)
Formalin is a chemical disinfectant for surface and premises beside its effect on several infective pathogens in ornamental and food fishes. Determination of formalin concentration for precious fish is critical. This problem is so important because of the aquarium fish varieties. In this study, OECD standard method was used for Koi fish formalin LC50 assessment. After a series of basic experime...