Comparison of pediatric radiation dose and vessel visibility on angiographic systems using piglets as a surrogate: antiscatter grid removal vs. lower detector air kerma settings with a grid — a preclinical investigation
نویسندگان
چکیده
The purpose of this study was to reduce pediatric doses while maintaining or improv-ing image quality scores without removing the grid from X-ray beam. This study was approved by the Institutional Animal Care and Use Committee. Three piglets (5, 14, and 20 kg) were imaged using six different selectable detector air kerma (Kair) per frame values (100%, 70%, 50%, 35%, 25%, 17.5%) with and without the grid. Number of distal branches visualized with diagnostic confidence relative to the injected vessel defined image quality score. Five pediatric interventional radiologists evaluated all images. Image quality score and piglet Kair were statistically compared using analysis of variance and receiver operating curve analysis to define the preferred dose setting and use of grid for a visibility of 2nd and 3rd order vessel branches. Grid removal reduced both dose to subject and imaging quality by 26%. Third order branches could only be visualized with the grid present; 100% detector Kair was required for smallest pig, while 70% detector Kair was adequate for the two larger pigs. Second order branches could be visualized with grid at 17.5% detector Kair for all three pig sizes. Without the grid, 50%, 35%, and 35% detector Kair were required for smallest to largest pig, respectively. Grid removal reduces both dose and image quality score. Image quality scores can be maintained with less dose to subject with the grid in the beam as opposed to removed. Smaller anatomy requires more dose to the detector to achieve the same image quality score.
منابع مشابه
Image Quality and Entrance Surface Dose Evaluation of Lateral Cervical Spine: A Study Using Grid and Non-Grid Techniques
Introduction: The purpose of this study is to investigate the effects of grid and non-grid techniques in the lateral cervical spine radiography on image quality and entrance surface dose (ESD). Although image quality and radiation doses have been studied by researchers, there is still a dearth of information on image quality and patient dose with different techniques. Material and Methods: The...
متن کاملDose optimization in pediatric cardiac x-ray imaging.
PURPOSE The aim of this research was to explore x-ray beam parameters with intent to optimize pediatric x-ray settings in the cardiac catheterization laboratory. This study examined the effects of peak x-ray tube voltage (kVp) and of copper (Cu) x-ray beam filtration independently on the image quality to dose balance for pediatric patient sizes. The impact of antiscatter grid removal on the ima...
متن کاملDesign and Manipulation 3D Imaging System by using Photodiode Grid
Introduction: Radiation imaging is one of the applicable methods in diagnostic medicine and nondestructive testing for industrial applications. In nondestructive 3D imaging, in addition to the radiation source, there is a requirement for a suitable detection system, data acquisition system, mechanical sections for moving objects, reconstruction algorithm and finally a computer for processing an...
متن کاملPatient dose values during interventional cardiology examinations in Yazd hospital, Iran
Background: The number of interventional cardiology (IC) procedures has increased rapidly. coronary angiography (CA) and percutaneous transluminal coronary angioplasty (PTCA) are now widely performed as a matter of routine, and they are considered safe procedures for experienced cardiologists. However, it is also known that these procedures are associated with high radiation doses due to long f...
متن کاملGrid: A location dependent intensity modulated radiotherapy for bulky tumors
Background: An effective method for management of large and bulky malignant tumors. This treatment modality is also known as Specially Fractionated Radiation Therapy (SFRT). In this treatment technique a grid block converted the open radiation field into a series of pencil beams. Dosimetric characteristics of an external beam grid radiation field have been investigated using experimental and Mo...
متن کامل