Wide tourniquet cuffs more effective at lower inflation pressures.
نویسندگان
چکیده
Longitudinal and radial tissue-fluid pressure distributions were determined beneath and adjacent to wide (12 and 18 cm) pneumatic tourniquet cuffs placed on intact human cadaveric arms and legs, respectively. Tissue fluid pressures exhibited relatively broad maxima at midcuff, and in most cases showed no differences at the various depths studied. Limb circumference was not a determining factor in the transmission of pressure to deeper tissue. We also investigated the effect of four cuff sizes (4.5, 8, 12, and 18 cm) on eliminating blood flow to the lower legs of normal subjects. The cuff pressure required to eliminate blood flow decreased as cuff width increased; thigh circumference was a determining factor in the pressure required to eliminate blood flow while using the smaller cuffs, but not while using the 18-cm cuff. Thus, wide cuffs transmit a greater percentage of the applied tourniquet pressure to deeper tissues than conventional cuffs; accordingly, lower cuff pressures are required, which may minimize soft-tissue damage during extremity surgery.
منابع مشابه
Occlusion of arterial flow in the extremities at subsystolic pressures through the use of wide tourniquet cuffs.
Tourniquet-induced peripheral neuropathy is at least partially attributable to excessive forces applied to the nerves beneath cuffs inflated to high pressures. Lowering the inflation pressure to the minimum necessary to obtain an effective arrest of blood flow distal to the tourniquet cuff should increase the safety of these systems. Tourniquet cuffs with widths varying from 4.5 cm to 80 cm wer...
متن کاملThe use of lower tourniquet inflation pressures in extremity surgery facilitated by curved and wide tourniquets and an integrated cuff inflation system.
Use of the lowest possible cuff inflation pressure should minimize the pathogenic effects of compression beneath the pneumatic tourniquet. Curved tourniquets (designed to fit conically shaped limbs) and wider tourniquets were associated with significantly lower arterial occlusion pressure (AOP) than standard, straight tourniquets on the arms and legs of 26 normal volunteers. These tourniquets w...
متن کاملTourniquet pain in a volunteer study: effect of changes in cuff width and pressure.
This study examines the relationship between pneumatic tourniquet cuff size, occlusion pressure and the resulting pain. Two tourniquet cuff widths were used, a wide (14 cm) and a narrow cuff (7 cm). Twenty volunteers were divided into two groups for tourniquet application: a pressure group in which the tourniquet was inflated to a pressure equal to the systolic pressure + 100 mmHg, and a satura...
متن کاملEffect of different cuff widths on the motor nerve conduction of the median nerve: an experimental study
BACKGROUND A bloodless operative field is considered mandatory for most surgical procedures on the upper and lower extremity. This is accomplished by using either an Esmarch bandage or a pneumatic tourniquet, but a number of complications are associated with both. Nerve palsy is one of the most frequently encountered complications of this procedure. Wider cuffs have been found to cause reduced ...
متن کاملA safe and effective low pressure tourniquet. A prospective evaluation.
An unusually wide pneumatic tourniquet has recently become available; we provide a simple formula for its use with a low but effective inflation pressure. A prospective, randomised and controlled trial in 600 lower limb and 150 upper limb operations is reported. The wide tourniquet (12.5 cm) was as effective at low pressure as was a conventional tourniquet (9 cm) inflated to significantly highe...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Acta orthopaedica Scandinavica
دوره 59 4 شماره
صفحات -
تاریخ انتشار 1988