Ion recombination correction factors (Pion) for Varian TrueBeam high‐dose‐rate therapy beams

نویسندگان

  • Stephen F. Kry
  • Richard Popple
  • Andrea Molineu
  • David S. Followill
چکیده

Ion recombination is approximately corrected for in the Task Group 51 protocol by Pion, which is calculated by a two-voltage measurement. This measurement approach may be a poor estimate of the true recombination, particularly if Pion is large (greater than 1.05). Concern exists that Pion in high-dose-per-pulse beams, such as flattening filter free (FFF) beams, may be unacceptably high, rendering the two-voltage measurement technique inappropriate. Therefore, Pion was measured for flattened beams of 6, 10, 15, and 18 MV and for FFF beams of 6 and 10 MV. The values for the FFF beams were verified with 1/V versus 1/Q curves (Jaffé plots). Pion was also measured for electron beams of 6, 12, 16, 18, and 20 MeV on a traditional accelerator, as well as on the high-dose-rate Varian TrueBeam accelerator. The measurements were made at a range of depths and with PTW, NEL, and Exradin Farmer-type chambers. Consistent with the increased dose per pulse, Pion was higher for FFF beams than for flattening filter beams. However, for all beams, measurement locations, and chambers examined, Pion never exceeded 1.018. Additionally, Pion was always within 0.3% of the recombination calculated from the Jaffé plots. We conclude that ion recombination can be adequately accounted for in high-dose-rate FFF beams using Pion determined with the standard two-voltage technique.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Ion recombination corrections of ionization chambers in flattening filter‐free photon radiation

The flattening filter free (FFF) X-rays can provide much higher dose rate at the treatment target compared to the conventional flattened X-rays. However, the substantial increase of dose rate for FFF beams may affect the ion recombination correction factor, which is required for accurate measurements using ionization chambers in clinical dosimetry. The purpose of this work is to investigate the...

متن کامل

Flattening filter free beams from TrueBeam and Versa HD units: Evaluation of the parameters for quality assurance.

PURPOSE Flattening filter free (FFF) beams generated by medical linear accelerators are today clinically used for stereotactical and non-stereotactical radiotherapy treatments. Such beams differ from the standard flattened beams (FF) in the high dose rate and the profile shape peaked on the beam central axis. Definition of new parameters as unflatness and slope for FFF beams has been proposed b...

متن کامل

Choice of a Suitable Dosimeter for Photon Percentage Depth Dose Measurements in Flattening Filter-Free Beams

The International Atomic Energy Agency Technical Reports Series-398 code of practice for dosimetry recommends measuring photon percentage depth dose (PDD) curves with parallel-plate chambers. This code of practice was published before flattening filter-free (FFF) beams were widely used in clinical linear accelerators. The choice of detector for PDD measurements needs to be reassessed for FFF be...

متن کامل

Dosimetry limitations and pre-treatment dose profile correction for sliding window IMRT

Background: This work investigated the dosimetry limitations of the random and systematic uncertainties of sliding window (SW) intensity modulated radiation therapy (IMRT). Materials and Methods: A Varian 21EX linear accelerator, Pinnacle3 treatment planning system and radiographic film dosimetry was used. The limitations of the SW were studied using beam modulation ranging from 2 to 10...

متن کامل

Monte Carlo modeling of HD120 multileaf collimator on Varian TrueBeam linear accelerator for verification of 6X and 6X FFF VMAT SABR treatment plans

A Monte Carlo (MC) validation of the vendor-supplied Varian TrueBeam 6 MV flattened (6X) phase-space file and the first implementation of the Siebers-Keall MC MLC model as applied to the HD120 MLC (for 6X flat and 6X flattening filter-free (6X FFF) beams) are described. The MC model is validated in the context of VMAT patient-specific quality assurance. The Monte Carlo commissioning process inv...