Gated myocardial perfusion SPECT underestimates left ventricular volumes and shows high variability compared to cardiac magnetic resonance imaging -- a comparison of four different commercial automated software packages
نویسندگان
چکیده
BACKGROUND We sought to compare quantification of left ventricular volumes and ejection fraction by different gated myocardial perfusion SPECT (MPS) programs with each other and to magnetic resonance (MR) imaging. METHODS N = 100 patients with known or suspected coronary artery disease were examined at rest with 99 mTc-tetrofosmin gated MPS and cardiac MR imaging. Left ventricular end-diastolic volume (EDV), end-systolic volume (ESV), stroke volume (SV) and ejection fraction (EF) were obtained by analysing gated MPS data with four different programs: Quantitative Gated SPECT (QGS), GE MyoMetrix, Emory Cardiac Toolbox (ECTb) and Exini heart. RESULTS All programs showed a mean bias compared to MR imaging of approximately -30% for EDV (-22 to -34%, p < 0.001 for all), ESV (-12 to -37%, p < 0.001 for ECTb, p < 0.05 for Exini, p = ns for QGS and MyoMetrix) and SV (-21 to -41%, p < 0.001 for all). Mean bias +/- 2 SD for EF (% of EF) was -9 +/- 27% (p < 0.01), 6 +/- 29% (p = ns), 15 +/- 27% (p < 0.001) and 0 +/- 28% (p = ns) for QGS, ECTb, MyoMetrix, and Exini, respectively. CONCLUSIONS Gated MPS, systematically underestimates left ventricular volumes by approximately 30% and shows a high variability, especially for ESV. For EF, accuracy was better, with a mean bias between -15 and 6% of EF. It may be of value to take this into consideration when determining absolute values of LV volumes and EF in a clinical setting.
منابع مشابه
RESPONSE TO REVIEW regarding MS ID
Title: Gated myocardial perfusion SPECT underestimates left ventricular volumes and shows high variability compared to cardiac magnetic resonance imaging-a comparison of four different commercial automated software packages Authors:
متن کاملEvaluation of the potential impact of reconstruction method on dyssynchrony parameters derived by phase analysis of gated-SPECT MPI: Comparison of two quantitative software
Introduction: Gated SPECT myocardial perfusion scanning has new capabilities in addition to its main applications such as left ventricular dyssynchrony using phase analysis. Phase analysis has been investigated through various software including Emory Cardiac Toolbox (ECTb) and Quantitative Gated SPECT (QGS). The aim of this study is to evaluate the effect of reconstruction par...
متن کاملPerformance of myocardial perfusion imaging using multi-focus fan beam collimator with resolution recovery reconstruction in a comparison with conventional SPECT
Objective: IQSPECT is an advanced high-speed SPECT modality for performing myocardial perfusion imaging (MPI), which uses a multi-focus fan beam collimator with resolution recovery reconstruction. The aim of this study was to compare IQSPECT compared with conventional SPECT interms of performance based on standard clinical protocols. In addition, we examined the concordance between convention...
متن کاملDiastolic left ventricular functional indices: Comparison between gated myocardial perfusion SPECT and echocardiography
Introduction: Many patients have only diastolic dysfunction without any abnormality in systolic function. We compared cardiac diastolic parameters obtained by gated myocardial perfusion SPECT (Gated SPECT) with those evaluated by echocardiography. Methods: Forty-nine patients (aged 37-85 years, 19 males and 30 women) underwent Gated SPECT and echocardiogr...
متن کاملThe prognostic value of stress/rest gated myocardial perfusion SPECT in patients with known or suspected coronary artery disease
Introduction:Gated myocardial perfusion SPECT has diagnostic and prognostic values in coronary artery disease (CAD). We tried to determine prognostic values of the left ventricular perfusion & functional indices as well as transient left ventricular dilation (TID) derived from gated myocardial perfusion SPECT. Methods: 1820patients who underwent gated myo...
متن کامل