Comparison of the McGrath® Series 5 and GlideScope® Ranger with the Macintosh laryngoscope by paramedics
نویسندگان
چکیده
BACKGROUND Out-of-hospital endotracheal intubation performed by paramedics using the Macintosh blade for direct laryngoscopy is associated with a high incidence of complications. The novel technique of video laryngoscopy has been shown to improve glottic view and intubation success in the operating room. The aim of this study was to compare glottic view, time of intubation and success rate of the McGrath® Series 5 and GlideScope® Ranger video laryngoscopes with the Macintosh laryngoscope by paramedics. METHODS Thirty paramedics performed six intubations in a randomised order with all three laryngoscopes in an airway simulator with a normal airway. Subsequently, every participant performed one intubation attempt with each device in the same manikin with simulated cervical spine rigidity using a cervical collar. Glottic view, time until visualisation of the glottis and time until first ventilation were evaluated. RESULTS Time until first ventilation was equivalent after three intubations in the first scenario. In the scenario with decreased cervical motion, the time until first ventilation was longer using the McGrath® compared to the GlideScope® and AMacintosh (p < 0.01). The success rate for endotracheal intubation was similar for all three devices. Glottic view was only improved using the McGrath® device (p < 0.001) compared to using the Macintosh blade. CONCLUSIONS The learning curve for video laryngoscopy in paramedics was steep in this study. However, these data do not support prehospital use of the McGrath® and GlideScope® devices by paramedics.
منابع مشابه
Comparison of the Incidence of Sore Throat and Hemodynamic Changes in Airway Management by Macintosh, Glideoscope Laryngoscope and Laryngeal Mask Airway Techniques: Randomized Controlled Trial
Background: Studies have shown that postoperative sore throat is the most common complication of tracheal intubation, However, the actual incidence and extent of it are not well documented. Objective: The aim of this study was to evaluate and compare the incidence of sore throat and hemodynamic Changes in Macintosh, Glidescope Laryngoscope and laryngeal mask airway techniques in elective surg...
متن کاملمقایسه میزان موفقیت لوله گذاری تراشه بعد از آموزش با دو روش گلایدسکوپ و لارنگوسکوپ
Background: Tracheal intubation is one of the most essential skills employed by all healthcare providers, nevertheless novice intubators turn out to be less successful when they initially perform the intubation. Video- laryngoscopes (GlideScopes) are novel instruments able to increase the success rate of tracheal intubation training. The aim of this study was to compare the success rate of trac...
متن کاملComparison of the Glidescope® and Pentax AWS® laryngoscopes to the Macintosh laryngoscope for use by Advanced Paramedics in easy and simulated difficult intubation
BACKGROUND Intubation of the trachea in the pre-hospital setting may be lifesaving in severely ill and injured patients. However, tracheal intubation is frequently difficult to perform in this challenging environment, is associated with a lower success rate, and failed tracheal intubation constitutes an important cause of morbidity. Novel indirect laryngoscopes, such as the Glidescope and the A...
متن کاملComparison of McGrath® Series 5 video laryngoscope with Macintosh laryngoscope: A prospective, randomised trial in patients with normal airways
OBJECTIVE The McGrath Video laryngoscope is a newly developed video laryngoscope that significantly improves laryngeal view and facilitates endotracheal intubation in difficult airways. However in patients with normal airway that is not mentioned before. The aim of this study was to compare the McGrath video laryngoscope with the Macintosh laryngoscope in patients with normal airways. METHODS...
متن کاملThe McGrath Series 5 video laryngoscope versus the Macintosh laryngoscope: a randomized trial in obstetric patients.
BACKGROUND/AIM Anesthesiologists have encountered various difficulties in securing the airway. Therefore, we compare the intubation times and hemodynamic changes between the McGrath Series 5 video laryngoscope and the Macintosh laryngoscope. MATERIALS AND METHODS A total of 80 obstetric patients were divided into 2 groups, orotracheally intubated with either the McGrath video laryngoscope or ...
متن کامل