Comparisons of the Tono-Pen® and Goldmann Applanation Tonometer in the Measurement of Intraocular Pressure of Primary Open Angle Glaucoma Patients in a Hospital Population in Southwest Nigeria

نویسندگان

  • Chizoba Onochie
  • Obiekwe Okoye
  • Adunola Ogunro
  • Tade Aribaba
  • Kunle Hassan
  • Adeola Onakoya
چکیده

OBJECTIVE The aim of this study was to compare intraocular pressure (IOP) measured with the Tono-Pen® to that measured using the Goldmann applanation tonometer (GAT) in patients with primary open angle glaucoma attending a glaucoma clinic. SUBJECTS AND METHODS A comparative clinic-based observational study was conducted involving 75 patients (39 men and 36 women) attending a glaucoma clinic in Southwest Nigeria. A pretested structured questionnaire was used to collect sociodemographic data. The Tono-Pen and the GAT were used to measure the IOP in each patient. Central corneal thickness (CCT) was measured with an ultrasonic pachymeter, and the corrected GAT value was calculated for each patient. The Tono-Pen readings were compared with the uncorrected and corrected GAT readings. Descriptive and comparative analyses were performed. Values for p < 0.05 were considered statistically significant. RESULTS The mean age of the participants was 60.39 ± 16.71 years. The mean IOP using the Tono-Pen was 21.1 ± 6.8 mm Hg versus the mean uncorrected GAT value of 17.1 ± 6.9 mm Hg and the corrected GAT value of 18.9 ± 7.5 mm Hg. The mean CCT was 510.5 ± 29.6 µm. The mean differences between the Tono-Pen reading and uncorrected and corrected GAT readings were 3.9 ± 2.6 and 2.1 ± 3.5 mm Hg, respectively. Gender (Tono-Pen vs. GAT; p = 0.981 vs. 0.437) and corneal thickness (p = 0.057) did not significantly affect the IOP value. Of the 75 patients, 68 (90.7%) preferred the Tono-Pen to the GAT. CONCLUSION In this study, the Tono-Pen gave a higher value for IOP than the uncorrected and corrected GAT values. Gender and corneal thickness did not significantly affect the measurements. Most patients found the Tono-Pen more acceptable than the GAT.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Intraocular Pressure Measurements by Three Different Tonometers in Children with Aphakic Glaucoma and a Thick Cornea

Background: To evaluate the agreement in intraocular pressure (IOP) measurements by Ocular Response Analyzer (ORA) and Tono-Pen XL (TXL) with the Goldmann Applanation Tonometer (GAT) and to examine corneal biomechanical properties in aphakic glaucoma patients with a central corneal thickness (CCT) >600 µ. Methods: Thirty-six eyes of aphakic glaucoma patients (group 1) and 40 eyes of normal chil...

متن کامل

Intraocular pressure measurement after penetrating keratoplasty: minified Goldmann applanation tonometer, pneumatonometer, and Tono-Pen versus manometry.

The accuracy of intraocular pressure measurement with the minified Goldmann applanation tonometer, the pneumatonometer, and the Tono-Pen tonometer were compared in post-mortem human eyes which had undergone penetrating keratoplasty. Enucleated post-mortem human eyes underwent same sized (7.75 mm) or 0.5 mm oversized (8.25 mm) autologous penetrating keratoplasty. Intraocular pressure was then se...

متن کامل

The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry.

AIMS To evaluate the influence of central corneal thickness (CCT) on intraocular pressure (IOP) measurements made with the Goldmann applanation tonometer (GAT), Tono-Pen XL, ocular blood flow tonograph (OBF), and Canon TX-10 non-contact tonometer (NCT). METHODS CCT was recorded for either eye (randomly selected) of each of 105 untreated patients with ocular hypertension and glaucoma attending...

متن کامل

Clinical comparison of the ProTon and Tono-Pen tonometers with the Goldmann applanation tonometer.

A clinical evaluation of a new electron ProTon tonometer was performed comparing the values of intraocular pressure (IOP) measured using this instrument with those determined by a similar instrument, Tono-Pen XL, and by Goldmann applanation tonometry. The mean IOP measured in 106 eyes with the ProTon tonometer was not significantly different from that determined with Goldmann applanation, while...

متن کامل

Effect of corneal thickness on intraocular pressure measurements with the pneumotonometer, Goldmann applanation tonometer, and Tono-Pen.

PURPOSE To compare intraocular pressure (IOP) measurements of the Ocular Blood Flow (OBF) pneumotonometer, Goldmann applanation tonometer, and Tono-Pen in eyes with normal corneas of various thicknesses. METHODS IOP was measured with an OBF pneumotonometer, Tono-Pen and Goldmann applanation tonometer in random order in 181 eyes with normal corneas. Central corneal thickness (CCT) was measured...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 25  شماره 

صفحات  -

تاریخ انتشار 2016