Rapid Deployment Aortic Valves Deliver Superior Hemodynamic Performance In Vitro

نویسندگان

  • Lisong Ai
  • Harvey Chen
  • Virginia Lin
  • Vinayak N. Bapat
چکیده

OBJECTIVE Clinical studies have demonstrated excellent hemodynamic performance of rapid deployment aortic valves; however, few studies have directly compared the performance of these valves with conventional bioprosthetic valves. Thus, the hemodynamic function of the EDWARDS INTUITY valve (rapid deployment valve) was compared with the Edwards Magna Ease valve in vitro (Edwards Lifesciences Corp, Irvine, CA USA). METHODS Elastomeric material was used to create an aortic root model that included a left ventricular outflow tract and aortic annulus. The model was based on reconstructions from 3-dimensional multislice computed tomography images in patients with aortic stenosis; the aortic root was scaled to a 21-mm effective annulus diameter. EDWARDS INTUITY valves (21-mm diameter) were deployed by stent frame expansion within the aortic root; Edwards Magna Ease valves (21-mm diameter) were sutured to the annulus. The left ventricular outflow tract area index (left ventricular outflow tract area/baseline area) and ellipticity or noncircularity as indexed by Dmax/Dmin were measured under a video microscope after valve placement. Hemodynamic data were collected under pulsatile flow with saline (70 beats per minute, 5 L/min, 100 mm Hg aortic pressure). RESULTS Compared with the Edwards Magna Ease valve (n = 4), the EDWARDS INTUITY valve (n = 4) had a greater effective orifice area (1.56 ± 0.01 vs 1.85 ± 0.06 cm, P < 0.001) and a lower transvalvular pressure gradient (23.4 ± 0.51 vs 16.8 ± 1.3 mm Hg, P < 0.001). Multiple regression analysis showed that 93% of the variation in the effective orifice area and transvalvular pressure gradient was due to variation in the left ventricular outflow tract area index and ellipticity index. CONCLUSIONS A clinically relevant aortic root model was developed to evaluate aortic valve performance. The superior performance of the EDWARDS INTUITY valve seemed to be related to both a greater inflow area and a more circular left ventricular outflow tract.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparative hydrodynamic evaluation of bioprosthetic heart valves.

BACKGROUND AND AIM OF STUDY Pressure gradients across cardiac valve prostheses have been identified as one of the most important performance measures in valve replacement surgery. Specifically in aortic valves, these gradients influence reduction of left ventricular hypertrophy and are postulated to influence long-term survival. The correct choice of replacement valve is hampered by a lack of u...

متن کامل

Edwards Intuity Aortic Bioprosthesis in Patient with Bicuspid Aortic Valve

Bicuspid aortic valve (BAV) is generally considered to be a contraindication to sutureless aortic valve replacement (AVR). Implantation of the Edwards Intuity aortic bioprosthesis is an innovative approach associated with superior hemodynamic performance, significantly reduced myocardial ischaemia and cardiopulmonary bypass times and proves to be suitable for type 1 and 2 of bicuspid aortic val...

متن کامل

Pledget-Armed Sutures Affect the Haemodynamic Performance of Biologic Aortic Valve Substitutes: A Preliminary Experimental and Computational Study

Surgical aortic valve replacement is the most common procedure of choice for the treatment of severe aortic stenosis. Bioprosthetic valves are traditionally sewed-in the aortic root by means of pledget-armed sutures during open-heart surgery. Recently, novel bioprostheses which include a stent-based anchoring system have been introduced to allow rapid implantation, therefore reducing the durati...

متن کامل

International Expert Consensus on Sutureless and Rapid Deployment Valves in Aortic Valve Replacement Using Minimally Invasive Approaches

OBJECTIVE To define the benefit of sutureless and rapid deployment valves in current minimally invasive approaches in isolated aortic valve replacement. METHODS A panel of 28 international experts with expertise in both minimally invasive aortic valve replacement and rapid deployment valves was constituted. After thorough literature review, the experts rated evidence-based recommendations in ...

متن کامل

Sutureless aortic valve replacement.

The increasing incidence of aortic stenosis and greater co-morbidities and risk profiles of the contemporary patient population has driven the development of minimally invasive aortic valve surgery and percutaneous transcatheter aortic valve implantation (TAVI) techniques to reduce surgical trauma. Recent technological developments have led to an alternative minimally invasive option which avoi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 12  شماره 

صفحات  -

تاریخ انتشار 2017