Re: Effect of beam profile measurement on arc therapy plan quality assurance: A case study

نویسندگان

  • Leonard H. Kim
  • Badal Juneja
چکیده

To the editor: In our recent study on beam profile measurement and its effect on arc therapy plan quality assurance, we asked the question, but could not answer, why the reported effect was observed in only one of two described commissioning experiences, despite the use of similar equipment and methods. In the publication, we speculated the difference was due to the difference in machines being modeled (i.e., Varian TrueBeam vs. Elekta Infinity with Agility MLC.) Since then, another experience commissioning a Varian TrueBeam suggests another possible answer. We observed that smoothing our measured data resulted in a noticeably shallower profile compared to unsmoothed measurements as well as Varian’s “Representative Beam Data.” In essence, we had created a virtual “volume averaging” effect. This was unexpected, because we followed the same smoothing procedure described for the “Representative Beam Data” as well as that used by other institutions. Nevertheless, something about this particular combination of measurement equipment, acquisition technique, and post-processing resulted in a noticeably altered profile. It is possible that similar circumstances resulted in the effects reported in our publication. It is thus advisable always to remember to assess the consistency of one’s data at each step of the commissioning process.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effect of beam profile measurement on arc therapy plan quality assurance: a case study

We present an example when profile measurement and modeling of an Elekta Agility multileaf collimator (MLC) had a large effect specifically on arc therapy plan quality assurance (QA) results using ArcCheck. ArcCheck absolute dose measurements of these plans were systematically lower than planned by 3-10%. Failing QA results were seen even with unmodulated static and conformal arcs. Furthermore,...

متن کامل

Quality Assurance of LINAC by Analyzing the Profile of 6-MV and 10-MV Photon Beams Using Star Track Device

Introduction: According to the American Society of Radiation Oncology, all patients receive radiation therapy during their illness, where radiation is delivered by the medical linear accelerator (Linac). The aim of this study was to evaluate the quality assurance (QA) of the Linac in analyzing the used dose profile in the treatment of cancer tumors. Materi...

متن کامل

Time‐resolved beam symmetry measurement for VMAT commissioning and quality assurance

In volumetric-modulated arc therapy (VMAT) treatment delivery perfect beam symmetry is assumed by the planning system. This study aims to test this assumption and present a method of measuring time-resolved beam symmetry measurement during a VMAT delivery that includes extreme variations of dose rate and gantry speed. The Sun Nuclear IC Profiler in gantry mount was used to measure time-resolved...

متن کامل

Quantification and modelling of the dosimetric impact of the treatment couch in volumetric modulated arc therapy (VMAT)

Background: As the volumetric modulated arc therapy (VMAT) becoming a main role of treatment ways, the effect of couch top becomes more significant. It is imperative to re-evaluate the couches that previously may have been considered of no importance during early treatment techniques. The impact of couch top on radiation delivery was explored and the couch model was tested with the aim of reduc...

متن کامل

Clinical experience with planning, quality assurance, and delivery of burst‐mode modulated arc therapy

"Burst-mode" modulated arc therapy (hereafter referred to as "mARC") is a form of volumetric-modulated arc therapy characterized by variable gantry rotation speed, static MLCs while the radiation beam is on, and MLC repositioning while the beam is off. We present our clinical experience with the planning techniques and plan quality assurance measurements of mARC delivery. Clinical mARC plans fo...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 18  شماره 

صفحات  -

تاریخ انتشار 2017