Sensitive determination of mercury by a miniaturized spectrophotometer after in situ single-drop microextraction.
نویسندگان
چکیده
An in situ single-drop microextraction (SDME) method was developed for trace mercury determination by a miniaturized spectrophotometer, in which a simple and cheap light-emitting diode (LED) was employed as the light source, and a handheld charge coupled device (CCD) was served as the detector. A droplet of 0.006% dithizone-CCl(4) (m/v) was used as extraction phase and hanged on a rolled PTFE tube. LED light was adjusted carefully to pass through the centre of the droplet and the entrance slit of the CCD detector. The radiation intensities of 475 nm before and after SDME (I(0) and I(i)) were recorded for quantification. Under the optimum conditions, the system provided a linear range of 2-50 μg L(-1), with a correlation coefficient of 0.9983 and a limit of detection (3σ) of 0.2 μg L(-1). The enrichment factor was about 69. The present method showed the merits of high sensitivity, simplicity, rapidity, low reagent consumption and field analysis potential. Finally, this method was successfully applied for the determination of the total mercury in spiked tap water sample, spiked river water sample and certified reference material (GBW (E) 080393, simulated water).
منابع مشابه
Comparison of Directly Suspended Drop Microextraction with Dispersive Liquid-liquid Microextraction Method for Extraction of Doxepin in Water and Biological Samples Prior to UV-Vis Spectrophotometer
The present research describes the use of directly suspended drop microextraction and air-assisted dispersive liquid-liquid microextraction method coupled with UV-Vis spectrophotometer for the analysis of doxepin in water and biological samples. Several important parameters which influence the extraction efficiencies of these two microextraction methods, such as the kind of extracting solven...
متن کاملIonic Liquid-based Ultrasound-assisted In-situ Solvent Formation Microextraction and High-performance Liquid Chromatography for the Trace Determination of Polycyclic Aromatic Hydrocarbons in Environmental Water Samples
A green and efficient ionic liquid-based ultrasound-assisted in-situ solvent formationmicroextraction (IL-UA-ISFME) in combination with high-performance liquid chromatographyultravioletdetection (HPLC-UV) has been successfully developed for the trace determination offive selected polycyclic aromatic hydrocarbons (PAHs) in environmental water samples. In thismethod, a hydrophobic ionic liquid (1...
متن کاملDetermination of Cadmium by Electrothermal Atomic Absorption Spectrometry after its Separation and Preconcentration by Syringe to Syringe Dispersive Liquid Phase Microextraction-Solidified Floating Organic Drop
The application of syringe to syringe dispersive liquid phase microextraction-solidified floating organic drop was extended for the separation and preconcentration of trace amounts of cadmium ions from water and cereal samples. The extracted cadmium was quantified by electrothermal atomic absorption spectrometry. Factors affecting the complex formation as well as microextraction efficiency such...
متن کاملPreconcentration of acrylamide with dispersive liquid-liquid microextraction based on solidification of floating organic drop prior to determination by HPLC
A novel, rapid, simple and sensitive dispersive liquid-liquid microextraction method based on the solidification of floating organic drop (DLLME-SFO) combined with high performance liquid chromatography-ultra violet detection (HPLC-UV) was used to determine acrylamide in potato chips. The derivation of the acrylamide happened in the presence of KBr, KBrO3, H2SO4. Based on studies, 1-undecanol w...
متن کاملSpectrophotometric Determination of Formaldehyde in Seawater Samples after In-situ Derivatization and Dispersive Liquid-Liquid Microextraction
In this paper, a simple dispersive liquid-liquid microextraction for the extraction and pre-concentration of formaldehyde in seawater samples followed with spectrophotometric is proposed. Formaldehyde was derivatized in situ with acetyl acetone in the presence of ammonium acetate in a single step. Then it was collected into a mixture of ethanol (disperser solvent) and chloroform (extracting...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of hazardous materials
دوره 183 1-3 شماره
صفحات -
تاریخ انتشار 2010