Comparison of pre-bent titanium mesh versus polyethylene implants in patient specific orbital reconstructions
نویسندگان
چکیده
INTRODUCTION Computerized tomography DICOM file can be relatively easily transformed to a virtual 3D model. With the help of additional software we are able to create the mirrored model of an undamaged orbit and on this basis produce an individual implant for the patient Authors decided to apply implants with any thickness, which are authors own invention to obtain volumetric support and more stable orbital wall reconstruction outcome. Material of choice was ultra-high molecular weight polyethylene (UHMWPE). OBJECTIVE The aim of this study was to present and compare functional results of individual reconstructions of orbital wall using either titanium mesh or ultra-high molecular weight polyethylene. MATERIALS AND METHODS 57 consecutive patients affected by orbital wall fracture (46 males, 11 females, mean age 34±14 year) were treated in Department of Maxillofacial Surgery from 2010 to 2012. In the first group we used patient specific treatment by titanium mesh shaped on a 3D printed model of a mirrored intact orbit (37 orbits) or by individually manufactured UHMW-PE implantby CAM milling in second group (20 orbits). All of these patients were subjected to preoperative helical computerized tomography and consultation of an ophthalmologist (including binocular single vision loss test - BSVL). Further on, patients were operated under general anaesthesia using transconjuctival approach. BSVL was again evaluated post-operationally in 1 month and 6 months later. RESULTS Functional treatment results (BSVL) for both groups were similar in 1 month as well as 6 months post operational time. There was no statistically significant difference between these two groups. CONCLUSIONS This study of 6 months functional result assessment of pre-bent individual implants and CNC milled ultra-high molecular weight polyethylene of the orbital wall has shown it to be a predictable reconstruction method. Individually shaped UHMWPE seems to be as good as pre-bent titanium mesh.
منابع مشابه
Preformed vs intraoperative bending of titanium mesh for orbital reconstruction.
OBJECTIVE The most accurate orbital reconstructions result from an anatomic repair of the premorbid orbital architecture. Many different techniques and materials have been used; unfortunately, there is currently no optimal method. This study compares the use of preformed vs intraoperative bending of titanium mesh for orbital reconstruction in 2-wall orbital fractures. STUDY DESIGN Cadaver-bas...
متن کاملPatient specific implants (PSI) in reconstruction of orbital floor and wall fractures.
Fractures of the orbital wall and floor can be challenging due to the demanding three-dimensional anatomy and limited intraoperative overview. Misfitting implants and inaccurate surgical technique may lead to visual disturbance and unaesthetic results. A new approach using individually manufactured titanium implants (KLS Martin, Group, Germany) for daily routine is presented in the current pape...
متن کاملPrecision of posttraumatic primary orbital reconstruction using individually bent titanium mesh with and without navigation: a retrospective study
BACKGROUND The aim of orbital wall reconstruction is to reestablish anatomically exact orbital volumes to avoid long-term complications. Navigation could facilitate complex reconstructions. METHODS Quality of the orbital reconstruction (n=94) was measured based on (A) volume changes and (B) on 3D shape deviations compared to the unaffected side. Volume analysis included segmentation of the or...
متن کاملUse of a Titanium Mesh Plate with High Three-Dimensional Flexibility to Repair an Orbital Floor Fracture: Clinical Note
Orbital blowout fractures are common sequelae of facial trauma. For their treatment, it is now possible to make three-dimensional (3D) implants on the basis of computed tomography (CT) images, using rapid prototyping and to bend the implants preoperatively in the desired shape. However, in orbital fracture, the unique anatomy of the orbit makes the process of fitting and aligning such implants ...
متن کاملA Wrapping Method for Inserting Titanium Micro-Mesh Implants in the Reconstruction of Blowout Fractures.
Titanium micro-mesh implants are widely used in orbital wall reconstructions because they have several advantageous characteristics. However, the rough and irregular marginal spurs of the cut edges of the titanium mesh sheet impede the efficacious and minimally traumatic insertion of the implant, because these spurs may catch or hook the orbital soft tissue, skin, or conjunctiva during the inse...
متن کامل