Evaluation of the Ohmeda 3700 pulse oximeter.
نویسندگان
چکیده
Arterial oxygen saturation values (Sao2) from 60% to 98% were measured by the Ohmeda 3700 pulse oximeter with the three types of probe available and compared with values of oxygen saturation estimated from direct arterial sampling (arterial oxygen and carbon dioxide tensions and pH) on 65 occasions. The response time of the oximeter was measured after a sudden rise in inspired oxygen concentration. Artefact rejection was assessed by arterial compression proximal to the probe site, and by simultaneous recordings of overnight Sao2 on opposite hands. The ability to recreate patterns of oscillating Sao2 from the data stored in the oximeter was also investigated. With the best probe system the oximeter measured Sao2, relative to arterial values estimated from Pao2, with a mean (SD) difference of -0.4% (1.8%). The response time was comparable with those of previous oximeters. It was not possible to generate artefactual dips in excess of 2% Sao2, and the dual overnight recordings rarely showed even small dips on one tracing alone. The stored data can recreate oscillating Sao2 signals with wavelengths down to about 35 seconds, but not below. The Ohmeda 3700 pulse oximeter appears to be suitable for unattended overnight recordings of Sao2.
منابع مشابه
Reliability of six pulse oximeters in chronic obstructive pulmonary disease.
Six pulse oximeters with finger probes were studied in three groups of 17 hypoxemic patients with COPD aged 50 to 75 years. Transcutaneous arterial oxygen saturation (SpO2) was measured with the Nellcor N101 (oximeter 1a), the Ohmeda Biox III (oximeter lb), the Nellcor N200 (oximeter 2a), the Critikon Oxyshuttle (oximeter 2b), the Radiometer Oxi100 (oximeter 3a), and the Ohmeda Biox 3700 (oxime...
متن کاملAvoiding hyperoxemia during neonatal resuscitation: time to response of different SpO2 monitors
AIM To assess the time to obtain reliable oxygen saturation readings by different pulse oximeters during neonatal resuscitation in the delivery room or NICU. METHODS Prospective study comparing three different pulse oximeters: Masimo Radical-7 compared simultaneously with Ohmeda Biox 3700 or with Nellcor N395, in newborn infants who required resuscitation. Members of the research team placed ...
متن کاملComparison of four pulse oximeters: effects of venous occlusion and cold-induced peripheral vasoconstriction.
The ability of four pulse oximeters (the Ohmeda 3700, Nellcor N100 and N200 and the Datex Oscar) to detect hypoxaemia was determined in the presence of venous obstruction and cold-induced peripheral vasoconstriction. Significant increases in detection time for hypoxaemia were found in both cases. There were no significant differences in detection time between the instruments, although the Ohmed...
متن کاملAccuracy of pulse oximeters in estimating heart rate at rest and during exercise.
Pulse oximeters are being widely used for non-invasive, simultaneous assessment of haemoglobin oxygen saturation. They are reliable, accurate, relatively inexpensive and portable. Pulse oximeters are often used for estimating heart rate at rest and during exercise. However, at present the data available to validate their use as heart rate monitors are not sufficient. We evaluated the accuracy o...
متن کاملPerformance of three new-generation pulse oximeters during motion and low perfusion in volunteers.
STUDY OBJECTIVE To evaluate pulse oximeter performance during motion and induced low perfusion in volunteers. DESIGN Prospective volunteer study. SETTING Direct Observation unit. SUBJECTS 10 healthy adult volunteers. INTERVENTIONS Ten volunteers were monitored with three different pulse oximeters while they underwent desaturation to about 75% oxygen saturation (SpO(2)) and performed mac...