Comparison of a powered bone marrow biopsy device with a manual system: results of a prospective randomised controlled trial
نویسندگان
چکیده
The diagnostic and clinical usefulness of a powered bone marrow biopsy device (OnControl()) versus a standard manual device (TRAP Hospital System) was studied. Primary endpoints were biopsy quality and patient pain during the procedure. Fifty patients underwent a total of 60 procedures by three expert operators in a randomised stratified fashion. Baseline demographic and clinical parameters were similar in both groups. The usage of conscious sedation, dosage of lidocaine/pethidin was similar between groups. Biopsy quality was rated 'sufficient for diagnosis' in 24/30 in the control group and 25/30 in the powered group (p=0.74). Biopsy cylinder length, procedure time (from skin contact of the biopsy needle to placement of the biopsy cylinder in the formalin container) and patient reported pain during the procedure (T1), 15 min after the procedure (T2) and 3-5 days after the procedure (T3) there were comparable between groups. In the small subgroup of patients that did not receive conscious sedation (n=15; manual 6, powered 9) significantly lower median pain scores were observed with the powered system (median pain score 3 vs 7; p=0.015). Patients were satisfied with either device whether sedation was used (sedation: median 9 for both groups, range 3-10 (manual) and 0-10 (powered)) no sedation (median 8 (manual) vs 9 (powered)). In summary bone marrow biopsies taken with the manual or powered device produce similar technical and clinical results. If no conscious sedation is used, pain during the procedure appears to be lower with the powered system. The use of a powered system seems to be justified in selected patients.
منابع مشابه
A powered bone marrow biopsy system versus manual methods: a systematic review and meta-analysis of randomised trials
A significant number of bone marrow biopsies are performed annually. Outcomes of patient pain and sample size of biopsy are important issues as perception of patient pain is underestimated by clinicians and sample size assists in accurate diagnosis. Manual extraction of marrow biopsy is the main method used. Recently a powered system has been introduced. The objectives of this systematic review...
متن کاملUsing a Powered Bone Marrow Biopsy System Results in Shorter Procedures, Causes Less Residual Pain to Adult Patients, and Yields Larger Specimens
BACKGROUND In recent years, a battery-powered bone marrow biopsy system was developed and cleared by the U.S. Food and Drug Administration to allow health care providers to access the bone marrow space quickly and efficiently. A multicenter randomized clinical trial was designed for adult patients to determine if the powered device had advantages over traditional manually-inserted needles in re...
متن کاملPowered bone marrow biopsy procedures produce larger core specimens, with less pain, in less time than with standard manual devices
Bone marrow sampling remains essential in the evaluation of hematopoietic and many non-hematopoietic disorders. One common limitation to these procedures is the discomfort experienced by patients. To address whether a Powered biopsy system could reduce discomfort while providing equivalent or better results, we performed a randomized trial in adult volunteers. Twenty-six subjects underwent bila...
متن کاملComparison of Transplantation of Bone Marrow Stromal Cells (BMSC) and Stem Cell Mobilization by Granulocyte Colony Stimulating Factor after Traumatic Brain Injury in Rat
Background: Recent clinical studies of treating traumatic brain injury (TBI) with autologous adult stem cells led us to compare effect of intravenous injection of bone marrow mesenchymal stem cells (BMSC) and bone marrow hematopoietic stem cell mobilization, induced by granulocyte colony stimulating factor (G-CSF), in rats with a cortical compact device. Methods: Forty adult male Wistar rats w...
متن کاملThe OnControl bone marrow biopsy technique is superior to the standard manual technique for hematologists-in-training: a prospective, randomized comparison
The purpose of this study was to compare a novel bone marrow device with the standard marrow needle in a prospective, randomized study in a teaching hospital employing hematologists-in-training. The new device, the OnControl Bone Marrow (OBM) Biopsy System, utilizes a battery-powered drill to insert the needle. Fifty-four bone marrows (27 standard and 27 OBM) were performed by 11 fellows under ...
متن کامل