Enhanced dynamic wedge output factors for Varian 2300CD and the case for a reference database

نویسنده

  • Christopher F. Njeh
چکیده

Dose inhomogeneity in treatment planning can be compensated using physical wedges. Enhanced dynamic wedges (EDW) were introduced by Varian to overcome some of the shortcomings of physical wedges. The objectives of this study were to measure EDW output factors for 6 MV and 20 MV photon energies for a Varian 2300CD. Secondly, to review the literature in terms of published enhanced dynamic wedge output factors (EDWOF) for different Varian models and thereby add credence to the case of the validity of reference databases. The enhanced dynamic wedge output factors were measured for the Varian 2300CD for both 6MV and 20 MV photon energies. Twelve papers with published EDWOF for different Varian linac models were found in the literature. Comparing our results with the published mean, we found an excellent agreement for 6 MV EDWOF, with the percentage differences ranging from 0.01% to 0.57%, with a mean of 0.03%. The coefficient of variation of published EDWOF ranged from 0.17% to 0.85% and 0.1% to 0.9% for the for 6 MV and 18 MV photon energies, respectively. This paper provides the first published EDWOF for 20 MV photon energy. In addition, we have provided the first compendium of EDWOFs for different Varian linac models. The consistency of value across models and institution provide further support that a standard dataset of basic photon and electron dosimetry could be established as a guide for future commissioning, beam modeling, and quality assurance purposes.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Optimum reckoning of contra lateral breast dose using physical wedge and enhanced dynamic wedge in radiotherapy treatment planning system

Background: It was intended to investigate the effect of physical wedge (PW) and enhanced dynamic wedges (EDW) on contralateral breast dose during primary breast irradiation in radiotherapy treatment, using high energy photon beams. Materials and Methods: The Varian’s Clinac dual mode linear accelerator model 2100 C/D and Siemen’s Primus accelerators were used for radiation doses wi...

متن کامل

Determination of an Effective Wedge Angle by Combination of Two Arbitrary Universal Wedge Fields in Radiation Therapy of Cancer Patients with Megavoltage Photon Beams.

Introduction: Wedge filters are commonly used in radiation oncology for eliminating hot spots and creating a uniform dose distribution in optimizing isodose curves in the target volume for clinical aspects. These are some limited standard physical wedges (15°, 30°, 45°, 60°), or creating an arbitrary wedge angle, like motorized wedge or dynamic wedge, ... The new formulation is...

متن کامل

Comparison of dosimetric characteristics of physical and enhanced dynamic wedges.

BACKGROUND Wedge filters can be used as missing tissue compensators or wedge pairs to alter the shape of isodose curves so that two beams can be angled with a small hinge angle at a target volume without creating a hotspot. AIM In this study the dosimetric properties of Varian Enhanced Dynamic Wedge (EDW) and physical wedges (PW) were analyzed and compared. MATERIALS AND METHODS Ionometric ...

متن کامل

Verification of Varian enhanced dynamic wedge implementation in MasterPlan treatment planning system

This paper investigates the accuracy of the two available calculation algorithms of the Oncentra MasterPlan three-dimensional treatment planning system (TPS)-- the pencil beam method and collapsed-cone convolution--in modeling the Varian enhanced dynamic wedge (EDW). Measurements were carried out for a dual high energy (6-15 MV) Varian DHX-S linear accelerator using ionization chambers for beam...

متن کامل

Clinical Implementation of Non - Physical Wedges 1999 AAPM Refresher Course

A non-physical wedge generates a spatial dose distribution similar to that produced by a physical wedge without a physical filter in the photon beam. Instead, an exponential fluence profile is produced via motion of one of the collimating jaws. Proposed in the late 1970s, nonphysical wedges have been implemented on both Varian and Siemens’ accelerators as the Varian Dynamic Wedge (DW) and Sieme...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 16  شماره 

صفحات  -

تاریخ انتشار 2015