System for prostate brachytherapy and biopsy in a standard 1.5 T MRI scanner.
نویسندگان
چکیده
A technique for transperineal high-dose-rate (HDR) prostate brachytherapy and needle biopsy in a standard 1.5 T MRI scanner is demonstrated. In each of eight procedures (in four patients with intermediate to high risk localized prostate cancer), four MRI-guided transperineal prostate biopsies were obtained followed by placement of 14-15 hollow transperineal catheters for HDR brachytherapy. Mean needle-placement accuracy was 2.1 mm, 95% of needle-placement errors were less than 4.0 mm, and the maximum needle-placement error was 4.4 mm. In addition to guiding the placement of biopsy needles and brachytherapy catheters, MR images were also used for brachytherapy treatment planning and optimization. Because 1.5 T MR images are directly acquired during the interventional procedure, dependence on deformable registration is reduced and online image quality is maximized.
منابع مشابه
Transrectal prostate biopsy and fiducial marker placement in a standard 1.5T magnetic resonance imaging scanner.
PURPOSE We investigated the accuracy and feasibility of a system that provides transrectal needle access to the prostate concurrent with 1.5 Tesla MRI which previously has not been possible. MATERIALS AND METHODS In 5 patients with previously diagnosed prostate cancer, MRI guided intraprostatic placement of gold fiducial markers (4 procedures) and/or prostate biopsy (3 procedures) was perform...
متن کاملAdvancements in magnetic resonance-guided robotic interventions in the prostate.
Magnetic resonance imaging (MRI) provides more detailed anatomical images of the prostate compared with the transrectal ultrasound imaging. Therefore, for the purpose of intervention in the prostate gland, diagnostic or therapeutic, MRI guidance offers a possibility of more precise targeting that may be crucial to the success of prostate interventions. However, access within the scanner is limi...
متن کاملAuthor's response to reviews Title: MRS-guided HDR brachytherapy boost to the dominant intraprostatic lesion in high risk localised prostate cancer Authors:
Background No concerns ---------------------------------------Material and Methods Only sextant biopsy? Typical guidelines usually recommend more sampling of the prostate. Unfortunately the sextant biopsy of prostate was the standard approach at the Townsville Hospital at the time of the study. A statement explaining that was added to the “Material and Methods” section, “Patient” subsection. --...
متن کاملRobotic Assistant for Transperineal Prostate Interventions in 3T Closed MRI
Numerous studies have demonstrated the efficacy of image-guided needle-based therapy and biopsy in the management of prostate cancer. The accuracy of traditional prostate interventions performed using transrectal ultrasound (TRUS) is limited by image fidelity, needle template guides, needle deflection and tissue deformation. Magnetic Resonance Imaging (MRI) is an ideal modality for guiding and ...
متن کاملMRI of the prostate: potential role of robots
Every year, 220,000 new cases of prostate cancer are detected. Approximately 18% of these cases will be treated with brachytherapy [1,2]. Over 1.2 million prostate needle biopsies are carried out every year in the USA. Systematic transrectal ultrasound (TRUS)-guided biopsy is the most frequently used technique for detecting prostate cancer and is the most commonly used method for guiding brachy...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Magnetic resonance in medicine
دوره 52 3 شماره
صفحات -
تاریخ انتشار 2004