[123I]FP-CIT ENC-DAT normal database: the impact of the reconstruction and quantification methods

نویسندگان

  • Livia Tossici-Bolt
  • John C. Dickson
  • Terez Sera
  • Jan Booij
  • Susanne Asenbaun-Nan
  • Maria C. Bagnara
  • Thierry Vander Borght
  • Cathrine Jonsson
  • Robin de Nijs
  • Swen Hesse
  • Pierre M. Koulibaly
  • Umit O. Akdemir
  • Michel Koole
  • Klaus Tatsch
  • Andrea Varrone
چکیده

BACKGROUND [123I]FP-CIT is a well-established radiotracer for the diagnosis of dopaminergic degenerative disorders. The European Normal Control Database of DaTSCAN (ENC-DAT) of healthy controls has provided age and gender-specific reference values for the [123I]FP-CIT specific binding ratio (SBR) under optimised protocols for image acquisition and processing. Simpler reconstruction methods, however, are in use in many hospitals, often without implementation of attenuation and scatter corrections. This study investigates the impact on the reference values of simpler approaches using two quantifications methods, BRASS and Southampton, and explores the performance of the striatal phantom calibration in their harmonisation. RESULTS BRASS and Southampton databases comprising 123 ENC-DAT subjects, from gamma cameras with parallel collimators, were reconstructed using filtered back projection (FBP) and iterative reconstruction OSEM without corrections (IRNC) and compared against the recommended OSEM with corrections for attenuation and scatter and septal penetration (ACSC), before and after applying phantom calibration. Differences between databases were quantified using the percentage difference of their SBR in the dopamine transporter-rich striatum, with their significance determined by the paired t test with Bonferroni correction. Attenuation and scatter losses, measured from the percentage difference between IRNC and ACSC databases, were of the order of 47% for both BRASS and Southampton quantifications. Phantom corrections were able to recover most of these losses, but the SBRs remained significantly lower than the "true" values (p < 0.001). Calibration provided, in fact, "first order" camera-dependent corrections, but could not include "second order" subject-dependent effects, such as septal penetration from extra-cranial activity. As for the ACSC databases, phantom calibration was instrumental in compensating for partial volume losses in BRASS (~67%, p < 0.001), while for the Southampton method, inherently free from them, it brought no significant changes and solely corrected for residual inter-camera variability (-0.2%, p = 0.44). CONCLUSIONS The ENC-DAT reference values are significantly dependent on the reconstruction and quantification methods and phantom calibration, while reducing the major part of their differences, is unable to fully harmonize them. Clinical use of any normal database, therefore, requires consistency with the processing methodology. Caution must be exercised when comparing data from different centres, recognising that the SBR may represent an "index" rather than a "true" value.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The impact of reconstruction and scanner characterisation on the diagnostic capability of a normal database for [123I]FP-CIT SPECT imaging

BACKGROUND The use of a normal database for [123I]FP-CIT SPECT imaging has been found to be helpful for cases which are difficult to interpret by visual assessment alone, and to improve reproducibility in scan interpretation. The aim of this study was to assess whether the use of different tomographic reconstructions affects the performance of a normal [123I]FP-CIT SPECT database and also wheth...

متن کامل

Improvement of diagnostic accuracy of Parkinson's disease on I-123-ioflupane single photon emission computed tomography (123I FP-CIT SPECT) using new Japanese normal database

Objective(s): I-123-ioflupane single photon emission computed tomography (FP-CIT-SPECT) has been used to assess dopamine transporter (DAT) loss in Parkinson's disease. The specific binding ratio (SBR), a quantitative parameter of DAT density in the striatum, may be affected by differences in age, sex, and SPECT system. The purpose of this study was to evaluate the util...

متن کامل

Impact of Dopamine Transporter Scan in Parkinson’s Disease

1.1 Technique of nigrostriatal FP-CIT SPECT The density of the presynaptic dopamine reuptake transporters (DAT) can be quantified with [123I]FP-CIT SPECT. The tracer Fluorpropyl-Carbomethoxy-Iodophenyl-Tropan (FPCIT), marked by radioactive 123iodine [123I] (therefore the name [123I]FP-CIT), binds reversibly to the DAT on the presynaptic membrane. Before application of FP-CIT, the iodine uptake ...

متن کامل

No difference in striatal dopamine transporter availability between active smokers, ex-smokers and non-smokers using [123I]FP-CIT (DaTSCAN) and SPECT

BACKGROUND Mesolimbic and nigrostriatal dopaminergic pathways play important roles in both the rewarding and conditioning effects of drugs. The dopamine transporter (DAT) is of central importance in regulating dopaminergic neurotransmission and in particular in activating the striatal D2-like receptors. Molecular imaging studies of the relationship between DAT availability/dopamine synthesis ca...

متن کامل

Quantification of striatal dopamine transporters with 123I-FP-CIT SPECT is influenced by the selective serotonin reuptake inhibitor paroxetine: a double-blind, placebo-controlled, crossover study in healthy control subjects.

UNLABELLED Dopamine transporter (DAT) imaging with (123)I-FP-CIT ((123)I-N-omega-fluoropropyl-2beta-carbomethoxy-3beta-(4-iodophenyl)nortropane) SPECT is frequently used to detect loss of nigrostriatal cells in parkinsonism. Recent (123)I-beta-CIT ((123)I-2beta-carbomethoxy-3beta-(4-iodophenyl)tropane) studies have shown a significant increase in striatal-to-nonspecific beta-CIT binding ratios ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 4  شماره 

صفحات  -

تاریخ انتشار 2017