Commissioning of a PTW 34070 large‐area plane‐parallel ionization chamber for small field megavoltage photon dosimetry

نویسندگان

  • Tom Kupfer
  • Joerg Lehmann
  • Duncan J. Butler
  • Ganesan Ramanathan
  • Tracy E. Bailey
  • Rick D. Franich
چکیده

PURPOSE This study investigates a large-area plane-parallel ionization chamber (LAC) for measurements of dose-area product in water (DAPw ) in megavoltage (MV) photon fields. METHODS Uniformity of electrode separation of the LAC (PTW34070 Bragg Peak Chamber, sensitive volume diameter: 8.16 cm) was measured using high-resolution microCT. Signal dependence on angle α of beam incidence for square 6 MV fields of side length s = 20 cm and 1 cm was measured in air. Polarity and recombination effects were characterized in 6, 10, and 18 MV photons fields. To assess the lateral setup tolerance, scanned LAC profiles of a 1 × 1 cm2 field were acquired. A 6 MV calibration coefficient, ND,w,LAC , was determined in a field collimated by a 5 cm diameter stereotactic cone with known DAPw . Additional calibrations in 10 × 10 cm2 fields at 6, 10, and 18 MV were performed. RESULTS Electrode separation is uniform and agrees with specifications. Volume-averaging leads to a signal increase proportional to ~1/cos(α) in small fields. Correction factors for polarity and recombination range between 0.9986 to 0.9996 and 1.0007 to 1.0024, respectively. Off-axis displacement by up to 0.5 cm did not change the measured signal in a 1 × 1 cm2 field. ND,w,LAC was 163.7 mGy cm-2 nC-1 and differs by +3.0% from the coefficient derived in the 10 × 10 cm2 6 MV field. Response in 10 and 18 MV fields increased by 1.0% and 2.7% compared to 6 MV. CONCLUSIONS The LAC requires only small correction factors for DAPw measurements and shows little energy dependence. Lateral setup errors of 0.5 cm are tolerated in 1 × 1 cm2 fields, but beam incidence must be kept as close to normal as possible. Calibration in 10 × 10 fields is not recommended because of the LAC's over-response. The accuracy of relative point-dose measurements in the field's periphery is an important limiting factor for the accuracy of DAPw measurements.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams

Introduction For dose measurement in Megavoltage (MV) photon beams with ion chambers, the effect of volume occupied by the air cavity is not negligible. Therefore, the result of measurement should be corrected with a displacement perturbation correction factor (Pdis) or using an effective point of measurement (EPOM). The aim of this study is to calculate the EPOM for cylindrical ion chamber and...

متن کامل

Small photon field dosimetry using EBT2 Gafchromic film and Monte Carlo simulation

Background: Small photon fields are increasingly used in modern radiotherapy especially in intensity modulated radiation therapy (IMRT) and stereotactic radiosurgery (SRS) treatments. Accurate beam profile and central axis depth doses measurements of such beams are complicated due to the electron disequilibrium. Hence the EBT2 (external beam therapy) Gafchromic film was used for dosimetry of sm...

متن کامل

Some steps towards establishing a tertiary standard dosimetry laboratory at a radiotherapy department

Background: In order to deliver the precise dose to the target in radiotherapy, absorbed dose to water at the reference point should be assessed. When the calibration procedure is performed for a reference dosimeter in the 60Co beam of a Secondary Standard Dosimetry Laboratory (SSDL), the total uncertainty in absorbed dose to water (Dw) is estimated to be approximately 1.5%. This study attempts...

متن کامل

Dosimetric Evaluation of Linac Photon Small Fields using MAGIC Polymer Gels

Introduction: In radiotherapy, methods of treatment planning are becoming increasingly more complicated. This requires verification of the doses delivered to increasingly smaller and more precise regions. Radiotherapy techniques are continuously employing smaller and smaller field sizes to deliver tighter radiation doses with higher therapeutic ratios, generating interest among researchers to p...

متن کامل

Evaluation of dosimetry a performance parameter for a homemade farmer-type ionization chamber according to IEC60731

Introduction: Due to its simplicity, fast response and high reliability, ion chamber detectors are one of the most suitable dosimetry instruments for X and gamma rays. The dosimeter    of photon beams, which used in radiotherapy centers, requires high accuracy and should pass requirements of IEC 60731 standards. About 80 percent of the IEC 60731 standard tests such as stabil...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 18  شماره 

صفحات  -

تاریخ انتشار 2017