Comparison of robot-assisted and conventional total knee arthroplasty: a controlled cadaver study using multiparameter quantitative three-dimensional CT assessment of alignment.
نویسندگان
چکیده
INTRODUCTION A functional total knee replacement has to be well aligned, which implies that it should lie along the mechanical axis and in the correct axial and rotational planes. Incorrect alignment will lead to abnormal wear, early mechanical loosening, and patellofemoral problems. There has been increased interest of late in total knee arthroplasty with robotic assistance. This study was conducted to determine whether robot-assisted total knee arthroplasty is superior to the conventional surgical method with regard to the precision of implant positioning. MATERIALS AND METHODS Twenty knee replacements, comprising ten robot-assisted procedures and ten conventional operations, were performed on ten cadavers. Two experienced surgeons performed the surgeries. Both procedures on each cadaver were performed by the same surgeon. The choice of which procedure was to be performed first was randomized. Following implantation of the prosthesis, the mechanical axis deviation, femoral coronal angle, tibial coronal angle, femoral sagittal angle, tibial sagittal angle, and femoral rotational alignment were measured via 3D CT scanning. These variables were then compared with the preoperatively planned values. RESULTS In the knees that underwent robot-assisted surgery, the mechanical axis deviation ranged from -1.94° to 2.13° (mean: -0.21°), the femoral coronal angle from 88.08° to 90.99° (mean: 89.81°), the tibial coronal angle from 89.01° to 92.36° (mean: 90.42°), the tibial sagittal angle from 81.72° to 86.24° (mean: 83.20°), and the femoral rotational alignment from 0.02° to 1.15° (mean: 0.52°) in relation to the transepicondylar axis. In the knees that underwent conventional surgery, the mechanical axis deviation ranged from -3.19° to 3.84° (mean: -0.48°), the femoral coronal angle from 88.36° to 92.29° (mean: 90.50°), the tibial coronal angle from 88.15° to 91.51° (mean: 89.83°), the tibial sagittal angle from 80.06° to 87.34° (mean: 84.50°), and the femoral rotational alignment from 0.32° to 4.13° (mean: 2.76°) in relation to the transepicondylar axis. In the conventional knee replacement group, there were two instances of outliers outside the range of 3° varus/valgus for the mechanical axis deviation. The robot-assisted knee replacements showed significantly superior femoral rotational alignment results compared with conventional surgery (p = 0.006). There was no statistically significant difference between robot-assisted and conventional total knee arthroplasty with regard to the other variables. All the measurements showed high intra-observer and inter-observer reliability. CONCLUSION Robot-assisted total knee arthroplasty showed excellent precision in the sagittal and coronal planes of the 3D CT scan. In particular, the robot-assisted technique showed better accuracy in femoral rotational alignment compared to the conventional surgery, despite the fact that the surgeons who performed the operations were more experienced and familiar with the conventional method than with robot-assisted surgery. It can thus be concluded that robot-assisted total knee arthroplasty is superior to conventional total knee arthroplasty.
منابع مشابه
Computer-assisted total knee replacement. A controlled cadaver study using a multi-parameter quantitative CT assessment of alignment (the Perth CT Protocol).
A controlled study, comparing computer- and conventional jig-assisted total knee replacement in six cadavers is presented. In order to provide a quantitative assessment of the alignment of the replacements, a CT-based technique which measures seven parameters of alignment has been devised and used. In this a multi-slice CT machine scanned in 2.5 mm slices from the acetabular roof to the dome of...
متن کاملComparison of Clinical, Functional, and Radiological Outcomes of Total Knee Arthroplasty Using Conventional and Patient-Specific Instrumentation
Background: Recently, patient-specific instrumentation (PSI) systems have been developed in order to increase theaccuracy of component positioning during total knee arthroplasty (TKA); however, the findings of previous studies arecontroversial in this regard. In the current randomized clinical study, the outcomes of computer tomography (CT)-basedPS (patient specific)-guided TK...
متن کاملEvaluation of Distal Femoral Rotational Alignment with Spiral CT Scan before Total Knee Arthroplasty (A Study in Iranian Population)
Background: Evaluating the landmarks for rotation of the distal femur is a challenge for orthopedic surgeons. Although the posterior femoral condyle axis is a good landmark for surgeons, the surgical transepicondylar axis may be a better option with the help of preoperative CT scanning. The purpose of this study was to ascertain relationships among the axes’ guiding distal femur rotational ...
متن کاملRobot-assisted implantation improves the precision of component position in minimally invasive TKA.
Minimally invasive and robot-assisted procedures have potential advantages when used for total knee arthroplasty (TKA). The purpose of this cadaveric study was to examine whether robot-assisted minimally invasive procedures improve TKA alignment after modifying the robotic techniques and instruments. Total knee arthroplasties were performed on 10 pairs of fresh cadaveric femora. Ten knees were ...
متن کاملUse of the Cusum technique for evaluation of a CT-based navigation system for total knee replacement.
Most of the early failures of total knee replacements are related to technical flaws. Conventional ancillary devices achieve good alignment in the frontal plane in only an average of 75% of total knee replacements. Computer-assisted surgery may improve the technical quality of implantation surgery. The aim of our study was to evaluate the use of computer-assisted surgery using a quality control...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Computer aided surgery : official journal of the International Society for Computer Aided Surgery
دوره 17 2 شماره
صفحات -
تاریخ انتشار 2012