The effect of scattered photons for x-ray air-kerma calibrations
نویسندگان
چکیده
The parallel-plate free-air ionization chambers are used for X-rays air-kerma rate standards at National Metrology Institute of Japan (NMIJ), National Institute of Advanced Industrial Science and Technology (AIST). The diaphragms are used to free-air chambers to fix the air volume of ionizing region in chambers. It means that absolute air-kerma rates are defined for collimated x-ray beams, because most of scattered photons are collimated by diaphragms. On the other hand, calibrated chambers are measured with non-collimated conditions. The variation of calibration constants was observed at different calibration distances and different diaphragm sizes for A6 Exradion chambers calibrations. The difference of the calibration constants are about 1.5% for 250kV X-ray beams. We thought that these results could be explained by the effect of scattered photons from the filters. In this study, we evaluate the contribution of scattered photons for free-air chambers and cavity chambers by Monte Carlo simulations, and suggest a new correction factor for evaluation of x-ray standards. The difference of calibration constants decreases below 0.4%.
منابع مشابه
Comparison Between the NIST and the KEBS for the Determination of Air Kerma Calibration Coefficients for Narrow X-Ray Spectra and 137Cs Gamma-Ray Beams
Air kerma calibration coefficients for a reference class ionization chamber from narrow x-ray spectra and cesium 137 gamma-ray beams were compared between the National Institute of Standards and Technology (NIST) and the Kenya Bureau of Standards (KEBS). A NIST reference-class transfer ionization chamber was calibrated by each laboratory in terms of the quantity air kerma in four x-ray referenc...
متن کاملAssessment of X-Ray Crosstalk in a Computed Tomography Scanner with Small Detector Elements Using Monte Carlo Method
Introduction: Crosstalk is a leakage of X-ray or light produced in a matrix of X-ray detectors or array of photodiodes in one element to other elements affecting on image contrast and spatial resolution. In this study, we assessed X-ray crosstalk in a computed tomography (CT) scanner with small detector elements to estimate the effect of various parameters such as X-ray tube voltage, detector e...
متن کاملConception and realization of a parallel-plate free-air ionization chamber for the absolute dosimetry of an ultrasoft X-ray beam.
We report the design of a millimeter-sized parallel plate free-air ionization chamber (IC) aimed at determining the absolute air kerma rate of an ultra-soft X-ray beam (E = 1.5 keV). The size of the IC was determined so that the measurement volume satisfies the condition of charged-particle equilibrium. The correction factors necessary to properly measure the absolute kerma using the IC have be...
متن کاملA comparison of shielding calculation methods for multi-slice computed tomography (CT) systems.
Currently in the UK, shielding calculations for computed tomography (CT) systems are based on the BIR-IPEM (British Institute of Radiology and Institute of Physics in Engineering in Medicine) working group publication from 2000. Concerns have been raised internationally regarding the accuracy of the dose plots on which this method depends and the effect that new scanner technologies may have. A...
متن کاملCalculation of the Scattered Radiation Profile in 64 Slice CT Scanners Using Experimental Measurement
Introduction: One of the most important parameters in x-ray CT imaging is the noise induced by detected scattered radiation. The detected scattered radiation is completely dependent on the scanner geometry as well as size, shape and material of the scanned object. The magnitude and spatial distribution of the scattered radiation in x-ray CT should be quantified for development of robust scatter...