Automatic correction of echo-planar imaging (EPI) ghosting artifacts in real-time interactive cardiac MRI using sensitivity encoding.

نویسندگان

  • Yoon-Chul Kim
  • Jon-Fredrik Nielsen
  • Krishna S Nayak
چکیده

PURPOSE To develop a method that automatically corrects ghosting artifacts due to echo-misalignment in interleaved gradient-echo echo-planar imaging (EPI) in arbitrary oblique or double-oblique scan planes. MATERIALS AND METHODS An automatic ghosting correction technique was developed based on an alternating EPI acquisition and the phased-array ghost elimination (PAGE) reconstruction method. The direction of k-space traversal is alternated at every temporal frame, enabling lower temporal-resolution ghost-free coil sensitivity maps to be dynamically estimated. The proposed method was compared with conventional one-dimensional (1D) phase correction in axial, oblique, and double-oblique scan planes in phantom and cardiac in vivo studies. The proposed method was also used in conjunction with two-fold acceleration. RESULTS The proposed method with nonaccelerated acquisition provided excellent suppression of ghosting artifacts in all scan planes, and was substantially more effective than conventional 1D phase correction in oblique and double-oblique scan planes. The feasibility of real-time reconstruction using the proposed technique was demonstrated in a scan protocol with 3.1-mm spatial and 60-msec temporal resolution. CONCLUSION The proposed technique with nonaccelerated acquisition provides excellent ghost suppression in arbitrary scan orientations without a calibration scan, and can be useful for real-time interactive imaging, in which scan planes are frequently changed with arbitrary oblique orientations.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Quantification and reduction of ghosting artifacts in interleaved echo-planar imaging.

A mathematical analysis of ghosting artifacts often seen in interleaved echo-planar images (EPI) is presented. These artifacts result from phase and amplitude discontinuities between lines of k-space in the phase-encoding direction, and timing misregistrations from system filter delays. Phase offsets and time delays are often measured using "reference" scans, to reduce ghosting through postproc...

متن کامل

Multi-echo segmented k-space imaging: an optimized hybrid sequence for ultrafast cardiac imaging.

Cardiac magnetic resonance imaging requires high temporal resolution to resolve motion and contrast uptake with low total scan times to avoid breathing artifacts. While spoiled gradient echo (SPGR) imaging is robust and reproducible, it is relatively inefficient and requires long breath-holds to acquire high time resolution movies of the heart. Echo planar imaging (EPI) is highly efficient with...

متن کامل

[Study of aliasing ERROR with SENSE in body diffusion image using single-shot EPI at 3T].

Magnetic field inhomogeneity causes artifacts in MRI. For example, in single-shot echo-planar imaging (EPI), they often appear as severe geometric distortions along the phase-encoding direction. Sensitivity encoding (SENSE) is useful in reducing the distortion in EPI since it only acquires partial k-space data using multiple receiver channels. In SENSE, a reference scan usually needs to be perf...

متن کامل

Reducing the respiratory motion artifacts in PET cardiology: A simulation study

  Introduction: There are several technical features that make PET an ideal device for the noninvasive evaluation of cardiac physiology. Organ motion due to respiration is a major challenge in diagnostic imaging, especially in cardiac PET imaging. These motions reduce image quality by spreading the radiotracer activity over an increased volume, distorting apparent les...

متن کامل

Monitoring, analysis, and correction of magnetic field fluctuations in echo planar imaging time series.

PURPOSE To assess the utility of concurrent magnetic field monitoring for observing and correcting for variations in k-space trajectories and global background fields that occur in single-shot echo planar imaging (EPI) time series as typically used in functional MRI (fMRI). METHODS Field monitoring was performed using an array of NMR field probes operated concurrently with series of single-sh...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of magnetic resonance imaging : JMRI

دوره 27 1  شماره 

صفحات  -

تاریخ انتشار 2008