Cloth destruction and haemolysis with totally cloth-covered Starr-Edwards prostheses.

نویسندگان

  • M Schottenfeld
  • J D Wisheart
  • J K Ross
  • J C Lincoln
  • D N Ross
چکیده

Four cases are described in which totally cloth-covered Starr-Edwards valves (model 2300) had to be removed. All were causing significant haemolysis, two in the absence of a peripheral leak. The principal operative finding was destruction of the Dacron covering the struts. Following replacement of these prostheses there was complete resolution of signs and symptoms. The possible causes of haemolysis and consequences of cloth destruction are discussed.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Thrombo-embolic complications of the cloth-covered Starr-Edwards prostheses No. 2300 aortic and No. 6300 mitral.

The thrombo-embolic complications of the cloth-covered Starr-Edwards prostheses No. 2300 aortic and No. 6300 mitral followed for an average of 14 months in 155 patients are reviewed. There was a high incidence of early fatal and disabling thrombo-embolus in patients 'having mitral valve replacement. Late emboli were more common after aortic valve replacement. Anticoagulant control was unsatisfa...

متن کامل

Haemolysis with Björk-Shiley and Starr-Edwards prosthetic heart valves: a comparative study.

Slater, S. D., Saflam, I. A., Bain, W. H., Turner, M. A., and Lawrie, T. D. V. (1974). Thorax, 29, 624-632. Haemolysis with Bjork-Shiley and Starr-Edwards prosthetic heart valves: a comparative study. A comparison was made of the haemolytic complications in 85 patients with two different types of Starr-Edwards cloth-covered ball and cage prosthesis with those in 44 patients with the Bjork-Shile...

متن کامل

Valve dysfunction of the cloth-covered Starr-Edwards ball valve.

BACKGROUND Between June 1968 and March 1977, Starr-Edwards cloth-covered ball valves were used for valve replacement on a routine basis. METHODS AND RESULTS Among the 66 operative survivors who underwent an isolated aortic or mitral valve replacement, 20 patients required reoperation 22 times because of valve dysfunction, thromboembolic complication, paravalvular leakage, hemolytic anemia, an...

متن کامل

Thirteen year-experience with mitral valve replacement.

At the Niigata University Hospital, various kinds of prosthetic valves such as the Starr-Edwards nonclothand cloth-covered, Bjork-Shiley tilting disc, and Carpentier-Edwards porcine valves, have been used from September, 1965 to March, 1979 for replacement of diseased valves on 176 patients. The patients with Starr-Edwards nonclothand cloth-covered ball valves showed survival rates of 90 and 84...

متن کامل

Results after mitral valve replacement with cloth-covered Starr-Edwards prostheses (models 6300, 6310/6320, and 6400).

The actuarial survival and thromboembolic rates for the three types of cloth-covered Starr-Edwards mitral prostheses, models 6300, 6310/6320, and 6400 followed 6, 5, and 2 years, respectively, were not significantly different throughout the years they were followed. The combined cumulative survival and thromboembolic proportion at 5 years for these prostheses were 71 and 66 per cent, respective...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Thorax

دوره 26 2  شماره 

صفحات  -

تاریخ انتشار 1971