The effect of a bellows leak in an Ohmeda 7810 ventilator on room contamination, inspired oxygen, airway pressure, and tidal volume.
نویسندگان
چکیده
We investigated the effect of a small bellows leak (bellows full at end-expiration) on inspired oxygen fraction (Fio(2)), exhaled tidal volume (Vt), airway pressure, and room contamination in an oxygen-driven anesthesia ventilator (Ohmeda 7810, Madison, WI). CO(2) concentration at the ventilator exhalation valve, Fio(2), Vt, and airway pressure were measured (n = 3) while ventilating a CO(2)-producing test lung at 8 breaths/min and an inspiratory/expiratory ratio of 1:2, with and without a bellows leak (4-mm-long tear). Set Vt was 400, 600, 800, and 1000 mL. Fresh gas flow (FGF) was 0.3 L/min O(2) and (a) 5.0 L/min air, (b) 2.0 L/min air, and (c) 0.2 L/min nitrogen. There was no clinical difference in Fio(2), Vt, PIP (peak inspiratory pressure) and PEEP (positive end-expiratory pressure), with and without a 4-mm bellows tear, at all FGFs and Vt settings. CO(2) at the ventilator exhalation valve was always nonzero with a bellows leak, indicating that CO(2)-laden circuit gas was contaminating the drive gas via the bellows leak. A 4-mm bellows tear in an Ohmeda 7810 ventilator allows anesthetic gases to contaminate ambient air but does not cause clinically significant changes in Fio(2), exhaled Vt, PIP, or PEEP.
منابع مشابه
Effect of Bubble and Ventilator-derived Continuous Positive Airway Pressure on the Management of Respiratory Distress Syndrome in Premature Neonates
Background: In this study, we aimed to compare ventilator-derived and bubble continuous positive airway pressure (CPAP) in neonates with respiratory distress syndrome admitted to Neonatal Intensive Care Unit of Vali-e-Asr Hospital, Birjand, Iran, in 2014. Methods: This cohort study was conducted among 68 patients assigned into two groups. The neonates in group A (32 infants) were treated with b...
متن کاملLow-tidal-volume ventilation in the acute respiratory distress syndrome.
Copyright © 2007 Massachusetts Medical Society. A 55-year-old man who is 178 cm tall and weighs 95 kg is hospitalized with community-acquired pneumonia and progressively severe dyspnea. His arterial oxygen saturation while breathing 100% oxygen through a face mask is 76%; a chest radiograph shows diffuse alveolar infiltrates with air bronchograms. He is intubated and receives mechanical ventila...
متن کاملA bench study of the effects of leak on ventilator performance during noninvasive ventilation.
BACKGROUND During noninvasive ventilation (NIV), leak around the mask may cause inadequate ventilatory support or patient-ventilator asynchrony such as auto-triggering. Some NIV ventilators may be better than others at compensating for leak. METHODS We bench studied 3 NIV ventilators (Vision, Carina, and Trilogy100) and 2 ICU ventilators (Puritan Bennett 840 and Evita XL) to assess how they ...
متن کاملStructure determines function
In this special issue of the Journal of Pediatric Intensive Care, we have compiled a series of articles relating to mechanical ventilation. We have chosen to select articles that do not address the usual reviews on mechanical ventilation but rather topics that our readers may find unique and interesting. The title of my editorial, “Structure determines function”, relates to that first lecture i...
متن کاملA comparison of the laryngeal tube with the laryngeal mask airway during routine surgical procedures.
UNLABELLED The laryngeal mask airway (LMA; Laryngeal Mask Company, Henley-on-Thames, UK) is an established airway device, whereas the laryngeal tube (LT) is relatively new and therefore not as well investigated. Therefore, the purpose of the present prospective, randomized, controlled trial was to compare the LT with the LMA in routine clinical practice. In 50 patients undergoing general anesth...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Anesthesia and analgesia
دوره 101 1 شماره
صفحات -
تاریخ انتشار 2005