Reduction of UH formation for U/U isotope ratio measurements at ultratrace level in double focusing sector field ICP-MS using D2O as solvent
نویسندگان
چکیده
The main factors affecting the accurate and precise determination of U using ICP-MS are instrumental background, the isobaric interference of UH molecular ion on U analyte ions, and the presence of U and U peak tails. An optimized analytical method for attenuating the influence of these factors on uranium isotope ratio measurements at ultratrace level of environmental samples has been developed. In order to reduce UH formation, D2O (heavy water) is used as a solvent for the dissolution and dilution of uranium samples. Abundance sensitivity was improved by use of medium mass resolution (m/Dm ~ 4450) in comparison with low mass resolution in double-focusing sector field ICP-MS (ICP-SFMS). For solution introduction the performances of several different sample introduction systems (Meinhard, Aridus and ultrasonic nebulizer) were studied. It has been shown, that for all nebulization systems, a diminution in UH/U is observed in D2O as compared with H2O as solvent. Optimum results were obtained in ICP-SFMS for a desolvating microconcentric nebulizer system (Aridus) with a minimum hydride formation rate of 9 6 10 and a limit for U/U isotopic ratio measurements of 3 – 5 6 10. A comparison was performed of three commercially available sector field ICP-MS devices, with good agreement found between single collector and multiple collector ICP-MS (MC-ICP-MS).
منابع مشابه
Isotopic analysis of uranium and plutonium using ICP-MS and estimation of burn-up of spent uranium in contaminated environmental samples{
Environmental monitoring of actinides and evaluation of the contamination source (nuclear weapons tests, nuclear power plant and nuclear reprocessing plant accidents, etc.) requires precise and accurate isotopic analysis of actinides, especially uranium and plutonium. Double-focusing sector-field inductively coupled plasma mass spectrometry (ICP-SFMS) using a low-flow microconcentric nebulizer ...
متن کاملDetermination of I/I isotope ratios in liquid solutions and environmental soil samples by ICP-MS with hexapole collision cell
The determination of I in environmental samples at ultratrace levels is very difficult by ICP-MS due to a high noise caused by Xe impurities in argon plasma gas (interference of Xe), possible IH2 1 interference and an insufficient abundance ratio sensitivity of the ICP mass spectrometer for I/I isotope ratio measurement. A sensitive, powerful and fast analytical technique for iodine isotope rat...
متن کاملUncertainty calculation on the determination of calcium absorption in nutritional experiments with enriched stable isotopes and detection by double focusing sector field ICP-MS with a shielded torch
A new method for the measurement of calcium isotope ratios and total calcium, using a double focusing sector field inductively coupled plasma mass spectrometry (ICP-MS) instrument equipped with a shielded torch, is presented. The method was applied for the measurement of the Ca/Ca, Ca/Ca and Ca/Ca ratios and total calcium by isotope dilution in urine samples from two different nutritional exper...
متن کاملRecent developments in isotope analysis by advanced mass spectrometric techniquesw Plenary lecture
An outstanding feature of mass spectrometry is the possibility of determining exact masses and accurate abundances of stable and unstable isotopes with high precision. The importance of isotope ratio measurements by mass spectrometry has grown in the last few years due to a significant improvement of instrumentation with respect to sensitivity, detection limits, precision and accuracy, whereby ...
متن کاملImaging mass spectrometry in biological tissues by laser ablation inductively coupled plasma mass spectrometry.
Of all the inorganic mass spectrometric techniques, laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) plays a key role as a powerful and sensitive microanalytical technique enabling multi- element trace analysis and isotope ratio measurements at trace and ultratrace level. LA-ICP-MS was used to produce images of detailed regionally-specific element distribution in 20 micro...
متن کامل