Acute changes in pacing threshold and R- or P-wave amplitude during permanent pacemaker implantation.

نویسندگان

  • M de Buitleir
  • W H Kou
  • S Schmaltz
  • F Morady
چکیده

This study examines the changes in pacing threshold and R- or P-wave amplitude during the first 30 minutes after implantation of tined and screw-in leads. The leads examined were those of 1 manufacturer (Medtronic) and consisted of 3 ventricular pacing leads (model numbers 6957 unipolar screw-in [11 patients], 6961 unipolar tined [12 patients] and 6962 bipolar tined [7 patients]) and 1 atrial lead (model number 6957J unipolar screw-in [10 patients]). After optimal lead position was obtained fluoroscopically in the right ventricular apex or right atrium, the pacing threshold and R- or P-wave amplitudes were measured at 5-minute intervals for 30 minutes. The acute ventricular pacing threshold with the screw-in lead was significantly higher than with the tined lead (0.84 +/- 0.17 vs 0.58 +/- 0.15 volts; p less than 0.001). There was a significant (p less than 0.001) acute decrease in the ventricular pacing threshold with both lead types, with the maximum decrease occurring 5 minutes after lead implantation. There was a significant acute increase in R-wave size with the ventricular screw-in lead that peaked 20 minutes after lead implantation (11.9 +/- 3.0 to 14.7 +/- 4.1 mV; p less than 0.001). The atrial screw-in lead behaved in a manner identical to its counterpart in the ventricle. In conclusion, there are acute changes in the pacing threshold and R- or P-wave amplitude obtained with tined and screw-in pacing leads.(ABSTRACT TRUNCATED AT 250 WORDS)

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Stable Atrial Sensing on Long-Term Follow Up of VDD Pacemakers

BACKGROUND The hemodynamic advantages of maintaining AV synchrony through AV synchronous pacing are widely known as compared to single chamber pacing. DDD pacemaker implantation entails higher cost and is technically more challenging than the VDD pacemaker. METHODS Seventy one patients underwent VDD lead (Biotronik GmbH, St. Jude Medical and Medtronic Inc.) implantation at KEM hospital, Mumba...

متن کامل

Acute results of permanent pacemaker implantation in Cipto Mangunkusumo General Hospital, Jakarta.

AIM To investigate the acute results (parameters at implant and clinical parameters) of permanent pacemaker implantation at our institution. METHODS Twenty five patients undergoing pacemaker implantation were included in this study. Subjects underwent medical history and functional class was assessed using New York Heart Association (NYHA) classification. All technical parameters (pacing thre...

متن کامل

Imaging in Cardiology: "Impossible" Pacemaker-Mediated Tachycardia during Implantation

A 78-year-old man was referred for pacemaker implantation due to symptomatic 3rd degree atrioventricular (AV) block. Neither the patient's history nor physical examination revealed abnormalities with regard to his cardiovascular status. Tachyarrhythmias such as intermittent atrial fibrillation or ventricular tachycardia had not been documented in recent years, and the patient was not on medicat...

متن کامل

[Foreseeable variation in parameters measured at implant and follow-up of permanent pacemaker active fixation electrodes].

OBJECTIVES To analyze the variations in the parameters relative to active fixation electrodes at the time of implantation and over subsequent follow-up during 6 months of the acute phase of implantation. DESIGN A descriptive, analytical, prospective, observational cohort study was made of consecutive cases over a period of 8 months (April-December 2010). SETTING Pacing unit of an Intensive ...

متن کامل

Periprocedural cardiac troponin elevation: a potential indicator of enhanced ventricular threshold in permanent pacemaker recipients

Background and aim: Permanent cardiac pacing is the most efficient treatment for patients with symptomatic bradycardia and high-degree atrioventricular (AV) block. For effective pacing, sensing and pacemaker battery longevity; ventricular pacing threshold (VPT) levels, lead impedance levels and Rwave amplitude levels must be desirable. We aimed to search the association between elevated serum t...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The American journal of cardiology

دوره 65 15  شماره 

صفحات  -

تاریخ انتشار 1990