Precision measurement of radioactivity in gamma-rays spectrometry using two HPGe detectors (BEGe-6530 and GC0818-7600SL models) comparison techniques: Application to the soil measurement

نویسندگان

  • Guembou Shouop Cebastien Joel
  • Samafou Penabei
  • Moyo Maurice Ndontchueng
  • Gregoire Chene
  • Eric Jilbert Nguelem Mekontso
  • Alexandre Ngwa Ebongue
  • Motapon Ousmanou
  • Strivay David
چکیده

To obtain high quality of results in gamma spectrometry, it is necessary to select the best HPGe detector for particular measurements, to calibrate energy and efficiency of gamma detector as accurate as possible. To achieve this aim, the convenient detector model and gamma source can be very useful. The purpose of this study was to evaluate the soil specific activity using two HPGe model (BEGe-6530 and GC0818-7600SL) by comparing the results of the two detectors and the technics used according to the detector type. The relative uncertainty activity concentration was calculated for 226Ra, 232Th and 40K. For broad energy germanium detector, BEGe-6530, the relative uncertainty concentration ranged from 2.85 to 3.09% with a mean of 2.99% for 226Ra, from 2.29 to 2.49% with a means of 2.36% for 232Th and from 3.47 to 22.37% with a mean of 12.52% for 40K. For GC0818-7600SL detector, it was ranged from 10.45 to 25.55% with a mean of 17.10% for 226Ra, from 2.54 to 3.56% with a means of 3.10% for 232Th and from 3.42 to 7.65% with a mean of 5.58% for 40K. The average report between GC0818-7600SL model and BEGe-6530 model was calculated and showed the mean value of 3.36. The main study was based on the following points: •Determination of The relative uncertainty activity concentration of 226Ra, 232Th and 40K•Determination of the relative uncertainty related to the radium equivalent activity to compare the performance of the two detection systems•Proved that the activity concentration determination in gamma spectrometry depended on the energy range emitted by a radionuclide. This study showed that the standard deviation measurement was less important to the result realized with BEGe-6530 HPGe model. Our findings were demonstrated that the results of the Broad Energy Germanium detector were more reliable.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A Comparative Study of Field Gamma-ray Spectrometry by NaI(Tl) and HPGe Detectors in the South Caspian Region

Natural radionuclides present in soil as well as certain anthropogenic radionuclides released to the environment are the major contributors to terrestrial outdoor exposures. In the assessment of human exposures from environmental radioactivity, besides the conventional method of soil and vegetation sampling combined with laboratory based analyses of environmental media, the other choice would b...

متن کامل

Measurement of Radioactivity Levels and Health Risks in the Surrounding Soil of Shazand Refinery Complex in Arak, Iran, Using Gamma-Ray Spectrometry Method

ntroduction: The purpose of this study was to measure the radioactivity in the agricultural soil of south-east of Shazand Refinery Complex to determine both reliable baseline data on the radiation level and the radiation dose exposure to the farmers and inhabitants of the studied area. Material and Methods: This study was conducted on 21 soil...

متن کامل

Optimization of measurement counting time in Gamma-ray spectrometry of soil samples

The method of optimal measurement counting time in Gamma spectrometry for soil samples was studied. Gamma spectrometry was done based on minimum detectable activity using the HPGe- p type with efficiency of 30% and FWHM 1.7keV (for 1.33 MeV 60Co). The samples were IAEA-RG Th-1, IAEA-RGU-1 and IAEA-RGK-1 prepared in bottles. The measurements were done for 12 different counting times from 5 min t...

متن کامل

Natural radioactivity concentration in raw materials used for manufacturing refractory products

Background: Some Particular areas contain natural radionuclide at levels much higher than those usually present in earth's crust. The radiological impact of the use of zirconium ore in zirconium industry is due to internal exposure of the lung by alpha particles and external exposure of the body by the gamma rays. The result of gamma spectrometry measurement of the concentrations of th...

متن کامل

Monitoring the concentration of 238U, 232Th, 40K and 137Cs of soil in Anarak-Khour district, Central Iran

This study was conducted to determine the radioactivity concentration in soil samples of Anarak-Khour district located at central Iran. 48 soil samples from a depth of 10-15 cm were taken from seven different sites for monitoring. The radioactivity concentration of 238U, 232Th, 137Cs and 40K in soil samples was determined, using a high resolution (HPGe detector, p-type) γ-spectrometry by detect...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 4  شماره 

صفحات  -

تاریخ انتشار 2017