Failure of BACTEC™ MGIT 960™ to detect Mycobacterium tuberculosis complex within a 42-day incubation period
نویسندگان
چکیده
For the optimal recovery of Mycobacterium tuberculosis from the BACTEC™ Mycobacterium Growth Indicator Tube 960™ system, an incubation period of 42-56 days is recommended by the manufacturer. Due to logistical reasons, it is common practice to follow an incubation period of 42 days. We undertook a retrospective study to document positive Mycobacterium Growth Indicator Tube cultures beyond the 42-day incubation period. In total, 98/110 (89%) were positive for M. tuberculosis complex. This alerted us to M. tuberculosis growth detection failure at 42 days.
منابع مشابه
Use of BACTEC MGIT 960 for recovery of mycobacteria from clinical specimens: multicenter study.
The BACTEC MGIT 960 instrument is a fully automated system that exploits the fluorescence of an oxygen sensor to detect growth of mycobacteria in culture. Its performance was compared to those of the radiometric BACTEC 460 instrument and egg-based Lowenstein-Jensen medium. An identical volume of sample was inoculated in different media, and incubation was carried out for 6 weeks with the automa...
متن کاملComparative evaluation of BACTEC MGIT 960 with BACTEC 9000 MB and LJ for isolation of mycobacteria in The Gambia.
BACKGROUND The BACTEC MGIT 960 was evaluated and compared with BACTEC 9000 MB and Lowenstein-Jensen medium for recovery rate of mycobacteria, time to detection, and contamination rate. METHODOLOGY 147 sputum samples obtained from patients with suspicion of tuberculosis were processed and inoculated into BACTEC MGIT 960, BACTEC 9000 MB and Lowenstein-Jensen medium using standardized procedures...
متن کاملEvaluation of BACTEC MGIT 960 system for the second-line drugs susceptibility testing of Mycobacterium tuberculosis in China.
When the performance of the BACTEC MGIT 960 system was evaluated with the traditional proportion method (PM) on 321 clinical isolates of Mycobacterium tuberculosis from China, concordance values were 98.44% capreomycin, 98.75% kanamycin, 98.75% ofloxacin and 94.08% ethionamide. The turnaround time with BACTEC MGIT 960 system (7.5±1.8 days) was significantly shorter than with PM (28 days or 42 d...
متن کاملMulticenter evaluation of the BACTEC MGIT 960 system for recovery of mycobacteria.
We evaluated the BACTEC MGIT 960 system, which is a fully automated, noninvasive system for the growth and detection of mycobacteria with a capacity to incubate and continuously monitor 960 7-ml culture tubes. We studied 3,330 specimens, 2,210 respiratory and 1,120 nonrespiratory specimens, collected from 2,346 patients treated at six sites. Processed specimens were inoculated into the BACTEC M...
متن کاملComparison of BACTEC MGIT 960 system and BACTEC 460 TB system for growth and detection of Mycobacteria from clinical specimens.
This study was carried out to compare the performance of BACTEC MGIT 960 with the BACTEC 460 TB for growth and detection of Mycobacteria from human clinical specimens. The BACTEC MGIT 960 instrument is a fully automated system that utilizes MGIT tubes containing an oxygen sensor embedded in silicon at the bottom and filled with 7 mL of modified Middlebrook 7H9 broth. Identical samples were inoc...