Automatic online embolus detection and artifact rejection with the first multifrequency transcranial Doppler.

نویسندگان

  • Rainer Brucher
  • David Russell
چکیده

BACKGROUND AND PURPOSE The goal of this study was to assess the first multifrequency transcranial Doppler system specially developed for online automatic detection of cerebral microemboli. METHODS The multifrequency Doppler instrumentation insonates simultaneously with 2.0- and 2.5-MHz frequencies. The detection threshold for embolus detection used in this study was a relative Doppler energy increase of >20 dB. ms, at which point the Doppler power increase was at least 5 dB and lasted >4 ms above the background energy. Four parameters were used in an optimized binary decision tree to recognize emboli: quarter Doppler shift, maximum duration limit, reference gate, and bidirectional enhancement. In in vitro studies, 200 plastic microspheres (80 micro m), 200 gas bubbles (8 to 25 micro m), and 600 artifacts were studied in a pulsatile closed-loop system. In vivo studies were carried out for 1 hour in 15 patients with mechanical heart valves and in 45 patients with carotid stenosis. This gave a total of 60 hours of online automatic monitoring in patients. RESULTS All 400 plastic spheres and microbubbles were automatically detected and correctly classified. Of the 600 artifacts, 596 (99.3%) were correctly classified as artifacts, and 4 (0.7%) were incorrectly identified as emboli (kappa=0.992, P<0.001). The experienced observer detected a total of 554 emboli and 800 artifacts in the heart valve (521 emboli, 400 artifacts) and carotid stenosis (33 emboli, 400 artifacts) patients. With multifrequency Doppler, 546 of these emboli (98.6%) and 791 of these artifacts (98.9%) were automatically detected and correctly classified as embolus or artifact (kappa=0.953, P<0.0001). CONCLUSIONS We found that multifrequency transcranial Doppler had a relatively high sensitivity and specificity when used to automatically detect cerebral microemboli and reject artifacts online.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Online automatic discrimination between solid and gaseous cerebral microemboli with the first multifrequency transcranial Doppler.

BACKGROUND AND PURPOSE The aim of this study was to assess the first multifrequency transcranial Doppler system that was specially developed to automatically detect and discriminate between solid and gaseous cerebral microemboli. METHODS The multifrequency transcranial Doppler instrumentation insonates simultaneously with 2.5 and 2.0 MHz. Differentiation between solid and gaseous microemboli ...

متن کامل

Embolus differentiation using multifrequency transcranial Doppler.

Embolus Differentiation Using Multifrequency Transcranial Doppler To the Editor: Markus and Punter report disappointing results for differentiation between solid and gaseous emboli using both the default settings of the Embo-Dop multifrequency transcranial Doppler system,1 and the modified criteria recommended by Russell and Brucher.2,3 One reason for this poor performance is a fundamental limi...

متن کامل

شناسایی میکرو آمبولی مغزی در سیگنال داپلر از روی ویژگی های غیر خطی

Abstract: An embolus is a blood clot, a fat globule or gas bubbles that may be freely circulating in bloodstream can stop the blood flow and lead to ischemia. In real time assessment of blood flow by Trans Cranial Doppler (TCD) method, travelling solid or gaseous micro emboli in the bloodstream by passing across the assessment area, causes a short time signal with high intensity. While TCD reco...

متن کامل

Automatic embolus detection by a neural network.

BACKGROUND AND PURPOSE Embolus detection using transcranial Doppler ultrasound is a useful method for the identification of active embolic sources in cerebrovascular diseases. Automated embolus detection systems have been developed to reduce the time of evaluation in long-term recordings and to provide more "objective" criteria. The purpose of this study was to evaluate the critical conditions ...

متن کامل

Computerized detection of cerebral emboli and discrimination from artifact using Doppler ultrasound.

BACKGROUND AND PURPOSE Transcranial Doppler ultrasound can detect circulating cerebral emboli. Monitoring of patients with potential embolic sources may allow identification of high-risk patients who can then be selected for prophylactic treatment. However, practical patient monitoring will require automated programs that can detect emboli and differentiate them from artifact. METHODS A new o...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Stroke

دوره 33 8  شماره 

صفحات  -

تاریخ انتشار 2002