Re-Analysis of the Uncertainty of the 0.895 µm Diameter (NIST SRM® 1690) and the 0.269 µm Diameter (NIST SRM® 1691) Sphere Standards
نویسنده
چکیده
The uncertainties of the mean diameters of the nominal 1.0 µm SRM(®) 1690 polystyrene spheres and of the nominal 0.3 µm SRM(®) 1691 polystyrene spheres are recomputed using the current NIST Guidelines for computing uncertainty. The revised expanded uncertainty (approximately 95 % confidence level) for SRM(®) 1690 polystyrene spheres is equal to 0.005 µm compared to previous value of 0.008 µm. The revised expanded uncertainty for SRM(®) 1691 is equal to 0.004 µm compared to the previous value of 0.007 µm. The major cause of the reduction in the uncertainty for the 1.0 µm spheres is from a decrease in the recomputed uncertainty of the refractive index of the polystyrene spheres. The 1.0 µm spheres were used in calibrating the electron microscope used to size the 0.3 µm spheres, and the reduction in the uncertainty of 1.0 µm SRM(®) uncertainty was the biggest factor in the decrease in the uncertainty of the 0.3 µm spheres.
منابع مشابه
Use of the Electrostatic Classification Method to Size 0.1 μm SRM Particles—A Feasibility Study
The use of the electrostatic classification method for sizing monodisperse 0.1 μm polystyrene latex (PSL) spheres has been investigated experimentally. The objective was to determine the feasibility of using electrostatic classification as a standard method of particle sizing in the development of a 0.1 μm particle diameter Standard Reference Material (SRM). The mean particle diameter was calcu...
متن کاملComparison of quartz standards for X-ray diffraction analysis: HSE A9950 (Sikron F600) and NIST SRM 1878.
A further comparison of the Health and Safety Executive (HSE) standard quartz, A9950 (Sikron F600), and the National Institute of Standards and Technology (NIST) Standard Reference Material (SRM) 1878, standard respirable alpha-quartz, has been carried out for the four principal diffraction peaks. In the earlier comparison by Jeyaratnam and Nagar (1993, Ann Occup Hyg; 37: 167-79), the standards...
متن کاملCertification of NIST SRM 1961: 30 μm Diameter Polystyrene Spheres
This report describes the certification of SRM 1961, an NIST Standard Reference Material for particle diameter. It consists of an aqueous suspension of monosize 30 μm diameter polystyrene spheres. The primary technique used optical microscopy; it gave a mean diameter value [Formula: see text] and a standard deviation of the size distribution σD = 0.21 μm. Over 2000 spheres were measured. The su...
متن کاملCertification of NIST SRM 1962: 3 μm Diameter Polystyrene Spheres
This report describes the certification of SRM 1962, a NIST Standard Reference Material for particle diameter. It consists of an aqueous suspension of monosize 3 (μm polystyrene spheres. Two calibration techniques were used: optical microscopy and electron microscopy. The first one gave a mean diameter of [Formula: see text] μm and a standard deviation of the size distribution σD = 0.020 μm, ba...
متن کاملCertification of NIST standard reference material 2389a, amino acids in 0.1 mol/L HCl--quantification by ID LC-MS/MS.
An isotope-dilution liquid chromatography-tandem mass spectrometry (ID LC-MS/MS) measurement procedure was developed to accurately quantify amino acid concentrations in National Institute of Standards and Technology (NIST) Standard Reference Material (SRM) 2389a-amino acids in 0.1 mol/L hydrochloric acid. Seventeen amino acids were quantified using selected reaction monitoring on a triple quadr...
متن کامل