Development of a Shield-Room Free Compact MRI System with a Highly Homogeneous RF Coil for Bone Density Measurements at the Calcaneus

نویسندگان

  • K. Taniguchi
  • S. Handa
  • K. Kose
چکیده

Introduction An important application of compact MRI is the measurement of trabecular bone volume fraction (TBVF) in peripheral bones for diagnosis of osteoporosis [1,2]. We reported a compact MRI system for TBVF measurement of the calcaneus and demonstrated its usefulness by discovering the age dependence of the relaxation times of the bone marrow protons [3]. However, there are still several disadvantages to this system. In this study, we developed a local RF shield to reduce the installation space, and an RF coil to produce a highly homogeneous RF magnetic field to reduce the measurement time by using a multiple spin echo (SE) sequence. Shield-Room Free Compact MRI System and Highly Homogeneous RF Coil The compact MRI system consists of a 0.2 T permanent magnet (160 mm vertical gap, 35 ppm homogeneity in 20 cm × 20 cm × 12 cm DEV), a gradient coil set, an RF probe, a local RF shield, and a compact MRI console. The local RF shield consists of an RF shield box, a flexible RF shield plane, and an LC balun circuit (Fig.1). The RF shield plane consists of a 0.1 mm thick copper sheet with thin plastic sheets attached to both sides. The RF coil is an oval solenoid (aperture: 175 mm × 85 mm, length: 120 mm) whose aperture is optimized for the calcaneus. As multiple SE sequences require a homogeneous RF magnetic field, the coil wire distribution of the oval solenoid was optimized using a genetic algorithm. Figures 2 and 3 show the RF coil and RF field distribution in the central plane measured using an oil phantom. Experiments To demonstrate the advantage of the homogeneous RF magnetic field, the right calcanei of 73 female subjects and six male subjects were measured using the following two protocols: two single SE sequences with different echo times (TE = 16, 96 ms) and a multiple SE (TE = 16, 32, 48, 64, 80, 96 ms) sequence. As the first protocol requires two separate scans, the measurement time of the first one (179.2 s) was twice that of the second (89.6 s). The calcanei of the six male subjects were measured 10 times repeatedly without repositioning to evaluate the reproducibility of the TBVF measurements. Results and Discussion Figure 4 shows the correlation between the TBVF measured for the 73 female subjects using the two protocols. The high correlation coefficient (R = 0.964, p < 0.0001) clearly shows equivalence of the two protocols. The percentage CV (coefficient of variance (standard deviation divided by the mean) expressed as a percentage) calculated for the repeated measurements averaged over the six male subjects was 2.2% and 2.9% for the multiple SE sequence and single SE sequences. This result clearly demonstrates the advantage of the multiple SE sequence over the single SE sequences in reproducibility. Although the application of single SE sequences to TBVF measurements has an advantage that the measured value is insensitive to the inhomogeneity of the RF magnetic field, the highly homogeneous RF field developed here reduced the measurement time by 50%, with a 30% CV reduction. To conclude, we achieved a reduction in installation space and measurement time in the compact MRI system for TBVF measurements of the calcaneus.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Development of a local electromagnetic shielding for an extremity magnetic resonance imaging system.

A local radio frequency (rf) shielding consisting of a Cu plate and an LC balun circuit has been developed for a compact magnetic resonance imaging (MRI) system with a 0.3 T permanent magnet. Performance of the local rf shielding was evaluated using an artificial external noise source irradiating a human subject whose hand was inserted into the rf coil of the MRI system. Power spectra of the rf...

متن کامل

Comparison of Bone Mineral Density Measured by Dual X-ray, Axial Dual-energy Photon X-ray Absorptiometry and Laser Absorptiometry of Calcaneus

The objective of this study was to validate the use of bone mineral measurements of the calcaneus bone by dual X-ray and laser (DXL) in a cross-sectional study carried out in an osteoporosis clinic. Measurements of bone mineral density (BMD) at proximal femur and spine were obtained by dual-energy photon X-ray absorptiometry (DEXA). Osteoporosis was defined by a DEXA T-score <-2.5 at the femora...

متن کامل

Assessment of the Characteristics of MRI Coils in Terms of RF Non-Homogeneity Using Routine Spin Echo Sequences

Introduction: One of the major causes of image non-uniformity in MRI is due to the existence of  non-homogeneity in RF receive and transmit. This can be the most effective source of error in  quantitative  studies  in  MRI  imaging.  Part  of  this  non-homogeneity  demonstrates  the  characteristics of RF coil and part of it is due to the interaction of RF field with the material being  imaged...

متن کامل

تعیین پروتکل بهینه جهت اندازه گیری میزان تراکم استخوان با استفاده از MRI Relaxometry و مقایسه نتایج حاصل با تکنیک DEXA

Introduction & Objective : The aim of this study was to optimize MRI conventional protocols for BMD(Bone Mineral Density) measurements using MRI Relaxometry in systems not facilitated with special multi echo protocols. Since, cortical and trabecular bone separation can not be performed in DEXA, so the results might lead to erroneous interpretation of BMD values. One method for bone quality dete...

متن کامل

Use of a radio frequency shield during 1.5 and 3.0 Tesla magnetic resonance imaging: experimental evaluation

Radiofrequency (RF) shields have been recently developed for the purpose of shielding portions of the patient's body during magnetic resonance imaging (MRI) examinations. We present an experimental evaluation of a commercially available RF shield in the MRI environment. All tests were performed on 1.5 T and 3.0 T clinical MRI scanners. The tests were repeated with and without the RF shield pres...

متن کامل