Simplified, totally enzymatic method for determination of serum triglycerides with a centrifugal analyzer.
نویسندگان
چکیده
We describe a totally enzymatic method for determination of serum triglycerides (triacylglycerols) specifically adaptable to the CentrifiChem system. The method involves lipolysis with lipase from Rhizopus arrhizus alone and quantitation of the resulting glycerol with glycerol dehydrogenase in a kinetic, fixed-time mode. Hydrolysis by the lipase is complete, for concentrations up to at least 5.0 g/liter, in 10 min at room temperature. The unfavorable equilibrium for the oxidation of glycerol is overcome by increasing the pH and adding excess NAD+. Under these conditions the glycerol determination is linear to at least 4.0 g of glycerol per liter, as triglyceride. The test exhibits acceptable accuracy and precision, and results correlate well with those by an alternative totally enzymatic procedure. The present method is unaffected by phosphatase and a considerably simplified reagent is used.
منابع مشابه
An integrated approach to lipid profiling: enzymatic determination of cholesterol and triglycerides with a centrifugal analyzer.
We describe enzymatic methods for determination of cholesterol and triglycerides (triacylglycerols), adapted for use with a centrifugal analyzer. Triglycerides are determined by a two-point kinetic method, which utilizes a single glycerol standard for calibration and provides for a within-run reagent blank correction. The endpoint determination of cholesterol combines the unique blanking capabi...
متن کاملManual and centrifugal analyzer determination of cholesterol and triglycerides with single-vial enzymatic reagents.
We have assessed Calbiochem single-vial enzymatic reagent systems for the analysis for cholesterol and triglycerides, using manual and centrifugal analyzer techniques. Analysis for cholesterol by both techniques had acceptable performance by the short-term criteria of the Center for Disease Control (CDC) during this study carried out under optimal conditions of variance. Analysis for triglyceri...
متن کاملRapid enzymatic determination of amylase in serum and urine using a centrifugal analyzer.
A convenient, totally enzymatic procedure for amylase assay using a centrifugal analyzer is described. The reaction scheme is examined using both starch and maltotetroase as substrates. Data for the reaction kinetics, linearity and sensitivity are presented. The proposed method exhibits linear reaction kinetics between three and five minutes after the initiation of the reaction and is linear to...
متن کاملImproved method for enzymatic determination of serum triglycerides.
I describe an enzymatic method for determing serum triglycerides (triacylglycerols). The triglycerides are hydrolyzed by a mixture of lipase and esterase. The glycerol released is determined by kinetic fixed-time analysis, with use of glycerol kinase, pyruvate kinase, and lactate dehydrogenase. Through addition of the competitive inhibitor ATP the Michaelis constant of pyruvate kinase is appare...
متن کاملAutomated enzymatic determination of L-lactate in serum, with use of a miniature centrifugal analyzer.
We describe an automated enzymatic reaction-rate method for spectrophotometric determination of lactate in serum with a miniature centrifugal analyzer. The L(+) -lactate is selectively oxidized in the presence of lactate dehydrogenase (EC 1.1.1.27) and NAD+ to form NADH, which is measured from its absorption. Reaction rates are determined automatically, and unknown concentrations are calculated...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 22 8 شماره
صفحات -
تاریخ انتشار 1976