Flow Injection Chemiluminescent Immunoassay for Carcinoembryonic Antigen Using Boronic Immunoaffinity Column
نویسندگان
چکیده
A flow injection chemiluminescence immunoassay for rapid and sensitive detection of carcinoembryonic antigen (CEA) by using a phenylboronic acid-based immunoaffinity column as a glycoprotein collector was proposed in this paper. The column was prepared by coupling of 3-aminophenylboronic acid on the glass beads through a γ-glycidoxypropyltrimethoxysilane (GPMS) linkage. Based on an indirect competitive immunoreaction, the mixture of CEA sample and enzyme conjugated CEA antibody (HRP-anti-CEA) was incubated in advance, followed by direct injection to the column to capture free HRP-labeled CEA antibody in the column. The trapped HRP-labeled antibody was detected by flow inject chemiluminescence in the presence of luminol and hydrogen peroxide. The decreased chemiluminescent signal was proportional to the concentration of CEA in the range of 3.0-30.0 ng/mL with a correlation coefficient of 0.998. The column showed an acceptable reproducibility and stability and is potentially used for practical clinical detection of the serum CEA level.
منابع مشابه
Flow injection immunoassay for carcinoembryonic antigen combined with time-resolved fluorometric detection.
Time-resolved fluorescence has been developed for immunoassay to obtain higher sensitivity than usual immunoassays. In this paper, a simple, sensitive and specific method was developed for immunoassay of serum carcinoembryonic antigen (CEA) by combining time-resolved fluoroimmunoassay (TRFIA) and flow injection analysis. Based on a sandwich immunoassay format, a monoclonal antibody immobilized ...
متن کاملNoncompetitive enzyme immunoassay for carcinoembryonic antigen by flow injection chemiluminescence.
BACKGROUND Recently, many automated immunoassay analyzers have been developed for carcinoembryonic antigen (CEA) to overcome the shortcomings in traditional immunoassay methods that are time-consuming and labor-intensive. Flow injection immunoassay (FIIA) has been increasingly applied to laboratory medicine due to its ease in automation, rapid speed and reproducible results. It is important to ...
متن کاملFlow-through multianalyte chemiluminescent immunosensing system with designed substrate zone-resolved technique for sequential detection of tumor markers.
A novel flow-through immunosensing system for performing a multianalyte chemiluminescent determination in a single run was designed. A new analytical strategy of substrate zone-resolved technique was proposed. Using carcinoma antigen 125 (CA 125) and carcinoembryonic antigen (CEA) as model analytes, the capture antibodies for CA 125 and CEA were immobilized on an UltraBind aldehyde-activated me...
متن کاملA channel-resolved approach coupled with magnet-captured technique for multianalyte chemiluminescent immunoassay.
A concept of channel-resolved multianalyte immunoassay (MAIA) and a semi-automated flow-through chemiluminescent (CL) MAIA system coupled with magnet-captured technique were proposed for rapid quantitation of different analytes in a single run. Using alpha-fetoprotein (AFP), carcinoembryonic antigen (CEA) and carcinoma antigen 125 (CA 125) as model analytes. They were firstly incubated in the m...
متن کاملChemiluminescent immunosensor for CA19-9 based on antigen immobilization on a cross-linked chitosan membrane.
A novel chemiluminescent immunosensor for carbohydrate antigen 19-9 (CA19-9) based on the immobilization of CA19-9 on the cross-linked chitosan membrane was developed. The different membranes were characterized by atomic force microscopy (AFM) and infrared spectrum, respectively. Based on a noncompetitive immunoassay format, this proposed chemiluminescent immunosensor enabled a low-cost, flexib...
متن کامل