Force‐Sensing Catheters During Pediatric Radiofrequency Ablation: The FEDERATION Study
نویسندگان
چکیده
BACKGROUND Based on data from studies of atrial fibrillation ablations, optimal parameters for the TactiCath (TC; St. Jude Medical, Inc) force-sensing ablation catheter are a contact force of 20 g and a force-time integral of 400 g·s for the creation of transmural lesions. We aimed to evaluate TC in pediatric and congenital heart disease patients undergoing ablation. METHODS AND RESULTS Comprehensive chart and case reviews were performed from June 2015 to March 2016. Of the 102 patients undergoing electrophysiology study plus ablation, 58 (57%) underwent ablation initially with a force-sensing catheter. Patients had an average age of 14 (2.4-23) years and weight of 58 (18-195) kg with 15 patients having abnormal cardiac anatomy. Electrophysiology diagnoses for the +TC group included 30 accessory pathway-mediated tachycardia, 24 atrioventricular nodal reentrant tachycardia, and 7 other. Baseline generator settings included a power of 20 W, temperature of 40°, and 6 cc/min flow during lesion creation with 11 patients (19%) having alterations to parameters. Seventeen patients (30%) converted to an alternate ablation source. A total of 516 lesions were performed using the TC with a median contact force of 6 g, force-time integral of 149 g·s, and lesion size index of 3.3. Median-term follow-up demonstrated 5 (10%) recurrences with no acute or median-term complications. CONCLUSIONS TactiCath can be effectively employed in the treatment of pediatric patients with congenital heart disease with lower forces than previously described in the atrial fibrillation literature. Patients with atrioventricular nodal reentrant tachycardia or atrioventricular reciprocating tachycardia may not require transmural lesions and the TC may provide surrogate markers for success during slow pathway ablation.
منابع مشابه
Contact Force and Atrial Fibrillation Ablation.
Catheters able to measure the force and vector of contact between the catheter tip and myocardium are now available. Pre-clinical work has established that the degree of contact between the radiofrequency ablation catheter and myocardium correlates with the size of the delivered lesion. Excess contact is associated with steam pops and perforation. Catheter contact varies within the left atrium ...
متن کاملUse of the new contact force sensing ablation catheter dramatically reduces fluoroscopy time during atrial fibrillation ablation procedures
OBJECTIVES To study the impact of contact force (CF) sensing on fluoroscopy, procedure, left atrial (LA) and ablation times and number of ablations during atrial fibrillation (AF) ablation. BACKGROUND Catheter ablation is an effective treatment for symptomatic AF. Recently a new ablation catheter providing real-time CF has been approved for use. METHODS A nested case-control study was perfo...
متن کاملTranscatheter ablation of supraventricular tachycardias in pediatric patients.
Ablation has become an important treatment for many pediatric patients with common supraventricular tachycardias (SVTs). Many multicenter studies have documented that radiofrequency (RF) catheter ablation is a safe and effective procedure for treatment of a large variety of SVTs in children and adults with a high success rate and minimal complications. Novel electrophysiology technologies such ...
متن کاملMRI-conditional catheter sensor for contact force and temperature monitoring during cardiac electrophysiological procedures
Background MR-guided cardiac electrophysiological (EP) ablations has drawn increasing attention from both the MRI and EP communities, as high-contrast MR images provide images that couple anatomical information with lesion efficacy [1]. Catheter manipulation can be challenging for cardiac electrophysiologists as conventional electroanatomical maps, frequently include false space. Perforation of...
متن کاملAn Investigation of Brightness Changes in Ultrasound Images due to Temperature Variations in Liver Tissue during Radiofrequency Ablation
Introduction: one of the minimally invasive methods of treatmenting liver malignancies is Radiofrequency Ablation (RFA) which can be applied to primary or secondary tumors. Nevertheless, the disease recurrence is probable after RFA application partially due to the inadequate capability of temperature monitoring of the target tissue and visualizing the thermal damage. The goal of this work was t...